Holding Field Control

From Hall A Wiki
Revision as of 19:46, 24 September 2008 by Jinhuang (Talk | contribs) (Vertical Coil)

Jump to: navigation, search

This is to discuss controling holding field of Polarized He3 Target with Helmholtz coils and correction coils.

Introduction

Due for Yi ...

Coil Current Control

Vertical Coil

Vertical Coil is controlled in constant current mode by Power supply 6675A. In addition, there is a DS345 (GPIB Address 14) providing a control voltage to 6675A. Vertical coil current is baseline current set in 6675A plus a shift proportion to the control voltage.

  • 6675A Baseline current is set by GPIB command CURR Base Line Current
  • DS345 control voltage is set by GPIB command OFFS Control Setting

A calibration is performed on Sept 23, 2008. With Base Line Current=15A and 16A. Error is on the 0.1% level.

File:Vertical Coil Curr vs Setting.png
Calibration Data: Data Points are marked by circles and linear fitting is solid line. Red marks data with Base Line Current=15A, while blue mark Base Line Current=16A
  • Base Line Current=15A Result:
     model:
       ''Control Setting''= a + b*''Coil Current''
     Coefficients (with 95% confidence bounds):
       a =      -2.652  (-2.654, -2.651)
       b =      0.1767  (0.1766, 0.1769)
  • Base Line Current=16A Result:
     model:
       ''Control Setting''= a + b*''Coil Current''
     Coefficients (with 95% confidence bounds):
       a =      -2.828  (-2.831, -2.826)
       b =      0.1767  (0.1765, 0.1769)
  • Further, it's reasonable to a extrapolation by assuming coefficient b in unrelated to Base Line Current:
       ''Control Setting''= 0.1767*(''Coil Current''-''Base Line Current'')

Horizontal Coils

Holding Field and Coil Current Relations

Field Roation GUI