Holding Field Control
This is to discuss controling holding field of Polarized He3 Target with Helmholtz coils and correction coils.
Contents
Introduction
Holding field is created by three orthognal Helmholtz coils. As shown here, the small and large coils are creating uniform magnetic field in horizontal directions and the vertical coil is for a vertical holding field. Since the fields from these three coils are orthognal to each other, a Coil System is defined.
In the configuration of Transveristy experiment, the coil system is rotated 37 degrees from the Hall System. That is small coil pointing to the right downstream of the beam and large coil to the left downstream.
Characteristics of Hoilding Field Coils
Coil Name | Radius (m) | Turns | Power Supply | I(A) for 35 G | U(V) for 35 G |
---|---|---|---|---|---|
Small | 0.667 | 256 | KEPCO | 10 | 35 |
Large | 0.758 | 272 | KEPCO | 10 | 35 |
Vertical | 0.946 | 203 | HP6675A | 18 | 49 |
Coil Current Control
Vertical Coil
Vertical Coil is controlled in constant current mode by Power supply 6675A. In addition, there is a DS345 (GPIB Address 14) providing a control voltage to 6675A. Vertical coil current is baseline current set in 6675A plus a shift proportion to the control voltage.
- 6675A Baseline current is set by GPIB command CURR Base Line Current
- DS345 control voltage is set by GPIB command OFFS Control Setting
A calibration is performed on Sept 23, 2008. With Base Line Current=15A and 16A. Error is on the 0.1% level.
- Base Line Current=15A Result:
model: ''Control Setting''= a + b*''Coil Current'' Coefficients (with 95% confidence bounds): a = -2.652 (-2.654, -2.651) b = 0.1767 (0.1766, 0.1769)
- Base Line Current=16A Result:
model: ''Control Setting''= a + b*''Coil Current'' Coefficients (with 95% confidence bounds): a = -2.828 (-2.831, -2.826) b = 0.1767 (0.1765, 0.1769)
- Further, it's reasonable to a extrapolation by assuming coefficient b in unrelated to Base Line Current:
''Control Setting''= 0.1767*(''Coil Current''-''Base Line Current'')
- According to vertical compass work which was done on Sep 27, to get vertical field, the currents are
I_v = 16 A I_s = -0.370 A I_l = 0.345 A
Horizontal Coils
Holding Field and Coil Current Relations
Gradient of holding field & Correction Coils
Based on our compass measurement, we exactly know the relationship between current and direction.
for longitudinal direction: 0.982[(0.056-0.424+I_l)/(0.101+0.291+I_s)]= tan(37.0508) for longitudinal direction: 0.982[(0.056-0.424+I_l)/(0.101+0.291+I_s)]= tan(127.0508) for vertical direction: General Equation is K[(I_v + E + BB)/sqrt{R*R(I_l+beta+delta)^2 +(I_s+ alpha+gamma)^2}] where K=0.546212 +/- 0.0022 and E= 0.176966 +/- 0.0354 for BigBite is off K= 0.552358 +/- 0.0022 and BB= 0.500829 +/- 0.034 for BigBite is on
On the other hand we need the amplitude of combined field to be about 25 Gauss, such as
I_l^2 + I_s^2 = 49 for in-plane case I_v = 14 A for vertical case
In each case, we have 2 equations and two variables. So we can determine the currents for each direction and take them into our field mapping data, choose the one of two settings which has less gradient. To choose the gradient made by correction coil, we use a code on matlab to minimize the eigenvalue of the gradient matrix. Then use following numerical formula to get the current in each correction coil.
for correction coils along small coil: I=-0.0342675967026 * dBz/dz for correction coils along large coil: I=-0.1327674624226 * dBx/dx
here and below we will use so-called coil-system, which is shown below :
longitudinal field with BigBite off
The current we use:
current in small coil: 5.1699340 A dBz/dz = -1.9151 current in large coil: 4.7192990 A dBx/dx = -14.1165
The gradient without correction coil: (in mG/cm)
dBx/dx dBy/dy dBz/dz dBy/dx dBz/dx dBz/dy average in pumping chamber: -6.593216 -6.993122 13.586337 -10.687164 4.486893 -0.722354 average in target chamber: -5.022424 -16.447674 21.470098 -0.880998 -6.230476 -1.056020
The currents of correction coils should be:
current in correction coils along small coil: 0.483739 A current in correction coils along large coil: 0.254263 A
transverse field with BigBite off
The current we use:
current in small coil: 4.0849622 A current in large coil: -5.6844599 A
The gradient without correction coil: (in mG/cm)
dBx/dx dBy/dy dBz/dz dBy/dx dBz/dx dBz/dy average in pumping chamber: 1.678019 -3.399559 1.721540 10.623624 -12.601686 -5.303687 average in target chamber: -4.639214 -10.318481 14.957695 0.601040 -14.587122 -1.811398
The current of correction coils should be:
current in correction coils along small coil: 0.26284 A dBz/dz = -7.6702 current in correction coils along large coil: 1.10504 A dBx/dx = -8.3231
longitudinal field with BigBite on
The current we use:
current in small coil: -3.7947230 A current in large coil: -5.8821831 A
The gradient without correction coil: (in mG/cm)
dBx/dx dBy/dy dBz/dz dBy/dx dBz/dx dBz/dy average in pumping chamber: 1.967711 -3.252200 1.284489 6.697551 -14.441990 3.611236 average in target chamber: -4.359505 -6.263064 10.622569 -1.467912 -19.770538 4.559035
The currents of correction coils should be:
current in correction coils along small coil: 0.41020711984781 A current in correction coils along large coil: 0.25867084703795 A
transverse field with BigBite off
The current we use:
current in small coil: -4.2409209 A current in large coil: 5.5690744 A
The gradient without correction coil: (in mG/cm)
dBx/dx dBy/dy dBz/dz dBy/dx dBz/dx dBz/dy average in pumping chamber: -5.901919 -7.173276 13.075196 -12.633972 0.479649 7.237270 average in target chamber: -4.727255 -12.117619 16.844874 -2.684100 -11.81095 4.831250
The current of correction coils should be:
current in correction coils along small coil: 0.52756678503488 A current in correction coils along large coil: 0.52210804597687 A