BCM calibration E06007

From Hall A Wiki
Revision as of 21:58, 3 February 2010 by Juaky (Talk | contribs)

Jump to: navigation, search

Wiki Entry

https://hallaweb.jlab.org/wiki/index.php/BCM_calibration_E04-018

PVDIS entry

http://hallaweb.jlab.org/experiment/E08-011/documents/bcm/AboutBCMcalibrations.html


I-scan runs in E06007 http://spreadsheets.google.com/ccc?key=0AomwfAy24yA3dEpFQ3k0QnhUQzJ2VEFKei0xaVdIY1E&hl=en


Carbon (1179-1183) Pb (1184-1188) Bi (1349-1354)
I (µA) T1 / T3 (Hz) I (µA) T1 / T3 (Hz) I (µA) T1 / T3 (Hz)
5.3 256 / 243 4.9 1.64k / 605 1.0 192 / 202
10 443 / 414 9.1 3.11k / 1.14k 2.1 420 / 420
21 884 / 826 19.5 6.64k / 2.43k 5 1.1k / 1.0k
40 1.77k / 1.65k 39.2 13.4k / 4.91k 10 2.0k / 2.0k
53 2.24k / 2.08k 42.8 14.7k / 5.40k 25 4.9k / 4.0k
40 8.0k / 7.2k



Steps for BCM calibration

Bcm e06007.gif


The Run faraday cup vs I0L02 is in [BcmLog125]

Faraday vs Injector current - From BCMLOg-125 - 20Feb07


Faraday_Current = (0.28+-0.03) + (0.994+-0.001) * IBC0L02_Current

Faraday / Injector = 1.0027 +- 0.0006 , for Current >5uA


Injector faraday calib 20Feb07.gif


Ratio Faraday vs I0L02 (All range) Ratio Faraday vs I0L02 (Current > 5uA)
Ratio injector faraday 20Feb07.gif Ratio injector faraday gt5uA 20Feb07.gif

Check of the Average Beam Current used in E06007 -- Only initial optics calibration runs used low currents -

A complete list of run information can be found in the link in the initial page: https://hallaweb.jlab.org/wiki/index.php/E06-007

BCM vs Run.gif


EPICS calibration from bcmlog-127 (Offsets without current in Hall from bcmlog-125)

Upstream Downstream
Offset 0.0009191 +/- 1e-07 0.00028730 +/- 5e-08
Constant 43.16 +/- 0.05 42.46 +/- 0.06
Constant (I>5uA) 43.25 +/- 0.02 42.60 +/- 0.02


EPICS CALIB.gif