Trigger Plane E & dE

From Hall A Wiki
Revision as of 05:06, 29 July 2010 by Rshneor (Talk | contribs)

(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

Energy loss calculation for protons in the E and dE detectors

E width 3 cm

dE width 0.3 cm


p=1000[MeV/c] T= 433[MeV] Eloss(E)= 9.3 [MeV] Eloss(dE)[MeV]=0.93

p=250[MeV/c] T= 33[MeV] Eloss(E)=33-5.7=27.3 [MeV] Eloss(dE)[MeV]=5.7

minimum ion Eloss(E)=6.6[MeV] Eloss(dE)=0.6[MeV]

protons punch-through point in the E plane [3 cm] at T=45 MeV = 294 [MeV/c]

commission guidelines -

we should try to set threshold for the E bars between 6.6[MeV] -> 9.3[MeV] = 8 [MeV]

max energy protons we would like to record is 45 MeV.

full range of adc is 4000

this gives -

45 MeV = 3000 ch

1 MeV = 66 ch

6 MeV = 530 ch (threshold)