HAND Gain calibration with known energy deposit, H(e,e'p)

From Hall A Wiki
Revision as of 11:18, 17 November 2010 by Igorko (Talk | contribs) (New page: == Energy deposit == Following script calculate the ADC channel for all PMTs in HAND. The Measured ADCs can be entered to the HV adjustment software in order to fine tune the desired Volt...)

(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

Energy deposit

Following script calculate the ADC channel for all PMTs in HAND. The Measured ADCs can be entered to the HV adjustment software in order to fine tune the desired Voltage.

In the script there is a cut that allows only events that seen by each bar and the bar that stand in the way of the optical line to the target.

The procedure is as following:

  1. calculate the desired votage using the gain curves
  2. make a measurement
  3. apply this script to extract the actual ADC
  4. correct the voltage, if needed, using the HV adjustment software