Difference between revisions of "Compton Photon Detector Centering Procedure"
From Hall A Wiki
m (→Spring 2019 Compton Photon Detector Commissioning) |
|||
Line 7: | Line 7: | ||
*# The following EPICS channels may be useful to monitor the backgrounds | *# The following EPICS channels may be useful to monitor the backgrounds | ||
*#* ComptonUSbg1 | *#* ComptonUSbg1 | ||
− | *#* ComptonUSbg2 <-- | + | *#* ComptonUSbg2 <-- May not be updating on EPICS/StripTools yet |
*#* ComptonDSbg1 | *#* ComptonDSbg1 | ||
*#* ComptonDSbg2 | *#* ComptonDSbg2 | ||
Line 54: | Line 54: | ||
*## Monitor EPICS channels: | *## Monitor EPICS channels: | ||
*##* ComptonVfingerRate <-- looking for a maximum/peak on this when scanning table position | *##* ComptonVfingerRate <-- looking for a maximum/peak on this when scanning table position | ||
− | *##* | + | *##* COM_DETPH_XCPOSai |
*##* ComptonCentralRate <-- possibly should see a minimum on this channel | *##* ComptonCentralRate <-- possibly should see a minimum on this channel | ||
*##* ComptonCentralRateHigh <-- possibly should see a minimum on this channel | *##* ComptonCentralRateHigh <-- possibly should see a minimum on this channel |
Revision as of 12:11, 24 June 2019
Spring 2019 Compton Photon Detector Commissioning
- Before MCC begins the initial process of passing bream through the Chicane, we must:
- Check and set HV:
- PhDet HV -> OFF
- Side (V) and Top (H) Finger Scintillator -> ON
- Laser table scintillators -> ON
- The following EPICS channels may be useful to monitor the backgrounds
- ComptonUSbg1
- ComptonUSbg2 <-- May not be updating on EPICS/StripTools yet
- ComptonDSbg1
- ComptonDSbg2
- The following EPICS values may also be useful
- hac_bcm_average
- IPM1P02A.XPOS
- IPM1P02A.YPOS
- IPM1P02B.XPOS
- IPM1P02B.YPOS
- Optionally the two finger scintillators may be useful, though it's likely they won't see anything if they are behind the collimator
- ComptonHfingerRate
- ComptonVfingerRate
- It may be useful to take a run during this period. Even though the HV on the detector is off, we may use it to study noise on the system.
- Check and set HV:
- After the beam is stable in the chicane and ready to center beam on laser
- Check and set HV:
- PhDet HV -> ON
- Side (V) and Top (H) Finger Scintillator -> ON
- Start a Compton Run (though, no need to analyze the run right now, this will just help us commission and calibrate the DAQ later)
- With the laser ON and 'not flipping, monitor the following EPICS channels:
- IPM1P02A.YPOS
- IPM1P02B.YPOS
- ComptonCentralRate
- ComptonCentralRateHigh
- COMPTON_PW1PCAV_ca
- As the beam is scanned vertically monitor for spikes in ComptonCentralRate and ComptonCentralRateHigh.
- The goal is to maximize ComptonCentralRate (assuming stable beam and laser power) and record the vertical position
- Ideally we would like the slope between BPM 02A and 02B to be as close to zero as possible
- Check and set HV:
- After the beam is centered on the laser, ask MCC to lock the beam in the optimal position found previously and then begin the detector centering procedure
- Check and make sure all HV Compton channels HV_Control_GUI are ON (should have been done by steps above, so just confirm nothing has tripped off)
- Start a new run
- Center the detector vertically (Note: do not lower the table to 0, always keep it above 0 mm)
- Set the Photon Detector table to X=69.00 mm, Y=6.00 mm
- Monitor EPICS channels:
- ComptonHfingerRate <-- looking for a maximum/peak on this when scanning table position
- COM_DETPH_YCPOSai
- ComptonCentralRate <-- possibly should see a minimum on this channel
- ComptonCentralRateHigh <-- possibly should see a minimum on this channel
- Then in 0.5 mm increments scan Y from 6.00 to 16.00 mm (~2-3 seconds at a time is enough)
- Once a rough idea of where the max was, scan a range in as small increments as possible with a +/- 2mm range
- Once you confirm the optimal y-position that gives maximum ComptonHfingerRate maximum (and minimum on ComptonCentralRate and ComptonCentralRate) , then move the table up by 20 mm from this position.
- Center the detector horizontally (Note: do not lower the table to 0, always keep it above 0 mm)
- Set the Photon Detector table to X=44.00 mm, Y=<value from step above>
- Monitor EPICS channels:
- ComptonVfingerRate <-- looking for a maximum/peak on this when scanning table position
- COM_DETPH_XCPOSai
- ComptonCentralRate <-- possibly should see a minimum on this channel
- ComptonCentralRateHigh <-- possibly should see a minimum on this channel
- Then in 0.5 mm increments scan Y from 44.00 to 54.00 mm (~2-3 seconds at a time is enough)
- Once a rough idea of where the max was, scan a range in as small increments as possible with a +/- 2mm range
- Once you confirm the optimal y-position that gives maximum ComptonVfingerRate maximum (and minimum on ComptonCentralRate and ComptonCentralRate) , then move the table in +x direction by 20 mm from this position.
- Center the detector horizontally (Note: do not lower the table to 0, always keep it above 0 mm)