Next: 5.5.3.10 Authorized Personnel
Previous: 5.5.3.6 Construction
Five photomultiplier tubes (PMT) are fixed of the two side walls. Each one
is surrounded by a high magnetic permeability shielding (mu-metal). The
fixing provides high voltage insulation between the PMT and the stell vessel.
A set of optic fibers provides light pulses to each PMT for their
calibration.
Characteristics
The type of the PMT is a Burle 8854 Photomultiplier. This is a 129 mm
(5 inch) diameter, 14 stage head on QUANTACON photomultiplier having
a bialkali photocathode of high quantum efficiency and an extremely
high gain cesiated gallium phosphide first
dynode followed by high stability copper beryllium dynodes in the succeeding
stages. The high gain first dynode permits the direct observation of peaks
corresponding to one two and three photoelectrons.
The 8854 features high quantum efficiency (22.5% at 385 nm),
ultraviolet response (UV response to 220 nm). The number of photon from the
Cerenkov light is proportionnal to so it's important to have a
extended response in UV light. Figure
show a typical curve for
the quantum efficiency as a function of the wavelenght.
Figure:
Quantum efficiency of the photocathode as a function of
the
|
Figure
shows a typical ADC pluse height spectrum
(with High Voltage around 2000V) where the single
photoelectron peak is very well separated from other sources of noise. Thus
they allow us to clearly define the electronic threshold to eliminate events
below this peak. The number of photoelectron produced in the Cerenkov is
expected to be large enough () to put a threshold around
few photoelectrons which garantee to get rid of the noise and have a very good
efficiency.
The same PMTs have been used for the aerogel Cerenkov detector for the TJNAF
Hall A and different configurations of the voltage dividers have been tested
(see ref NIM A309(1991)552-554). The optimized configuration with high
photocathode to first dynode voltage has been adopted.
Figure:
Typical ADC spectrum. The single peak is clearly
separated form the noise and the peak for 2 is also seen.
|
Numbering scheme
Figure
and figure
show the correspondance between
the serial number of the PMT and the number from 1 to 10 in the Cerenkov
(which correspond to a position in the box) in the hadron and electron arm.
Figure:
Numerotation for the PMT of the Cerenkov counter of the
electron arm.
|
Figure:
Numerotation for the PMT of the Cerenkov counter of the
hadron arm.
|
High Voltage adjustment
If an electron is above the Cerenkov threshold, it will produce Cerenkov
light in the gas. This light will be translate into an electric signal
via a photomultiplicator. Each photomutiplicator will receive the light
transmitted by his mirror. As a group of mirror can be touched by a good
event, we expect to have sometimes a group of touched PMT. Before the
discrimination of the signal we have to sum the signals over the touched
PMT. In order to be able to make the summ, we have to know a relative
calibration between the ten PMT of a counter.
For the further analysis of experiment
which will use these Cerenkov counters, we need to know the efficiency of
these counters. From the spectrum of charge, we can find this efficiency from
the number of mean photoelectrons in this spectrum (see formula
). To
obtain this mean number of photoelectrons from the ADC spectrum, we need to
make an absolute calibration in energy of each PMT.
The high Voltage has to be adjust in order to have for each PMT the position
of the photoelectron peak at the same position which is around few 20 channels
above the pedestal. This adjustment has to verify some conditions:
- -
- to correspond to a High Voltage where the PMT noise is
not high (below 2kHz)
- -
- to correspond to the maximum value of the quantum efficiency
of the photocathode.
If the two conditions are not possible at the same time, a compromise has
to be found.
Noise measurement
Figure
shows the PMT noise as a function of the High Voltage. The
analogic signal of each PMT was dicriminated with a low threshold (60 mV) and
the result of this discrimination was going to a scaler. The scaler rate
represent the noise in this case because it was done with the PMT in their
final position, in the Cerenkov box which has no leak of light, and without
beam or any source light except cosmic of course...
We can see from the measurement that below a High Voltage of 2500V, all PMT
have a noise rate below 2 kHz except the number 3 (in electron Cerenkov) which
has to be perhaps change for the future. This rate is relatively high because
the threshold is very low. As we are working with threshold Cerenkov counters
which give only photoelectrons (theoretical value which has to be
confirm with electron beam) we have to work with a low threshold.
Figure:
PMT noise as a function of the HV. This was obtained with a
low threshold of discrimination ( 60 mV)
|
Relative gain measurement
From figure
you can extract a gain (in arbitrary unit) from the
position of the single photoelectron peak. This was done for all the 20 PMT
(10 for each Cerenkov counter) at three different High Voltage, 1800, 2000
and 2200 Volts. The PMT was placed in a black box (no leak of light) and an
optical fiber was coming just in front of the photocathode. A source of
green light (diode) with adjustable amplitude was used and once
we suceed to tune this amplitude in order to see the photoelectron peak in a
given PMT, the same amplitude was used for each PMT.
The results are reported in the table
and are also
plotted in figurepmga.
Table:
Relative gains for all the PMT in arbitrary units.
PMT n0 |
HV1=1800 V |
HV2=2000 V |
HV3=2200 V |
1 |
20 |
75 |
231 |
2 |
45 |
130 |
390 |
3 |
31 |
123 |
saturation |
4 |
21 |
78 |
233 |
5 |
40 |
132 |
410 |
6 |
21 |
75 |
243 |
7 |
18 |
30 |
90 |
8 |
24 |
93 |
283 |
9 |
21 |
74 |
255 |
10 |
16 |
56 |
178 |
11 |
30 |
95 |
280 |
12 |
15 |
45 |
133 |
13 |
20 |
75 |
230 |
14 |
17 |
65 |
208 |
15 |
20 |
52 |
150 |
16 |
21 |
76 |
237 |
17 |
25 |
80 |
250 |
18 |
15 |
62 |
170 |
19 |
15 |
53 |
171 |
20 |
no peak |
15 |
51 |
We can see that number 3 has a relative gain higher than the others. The noise
measurement shows already that this PMT get a lot of
noise and at 2200 Volts it is on saturation. The PMT number 20 has a low
gain and no single photoelectron peak was seen at 1800 V.
Figure:
Relative gain in arbitrary units for all the PMTs at three
different High Voltage 1800,2000 and 2200 Volts.
|
Total radiation lenght
When a particle cross the Cerenkov detector, it will go through:
- -
- the entrance and exit window
- -
- the gas medium
- -
- one mirror
The total radiation lengh for the whole Cerenkov detector is
wich is equivalent of 5 mm of scintillator.
Next: 5.5.3.10 Authorized Personnel
Previous: 5.5.3.6 Construction
10/13/1997