Next: 5.3.3.18 Troubleshooting
Previous: 5.3.3.14 Safety Assessments
All the operations are handled by computers. Some of them can be also
performed independently by mechanical switches (referred as manually). In the
following, the main window of the slow control program on the MacA or MacB
display will be referred simply as the "main window".
Connection of MacB to MacA
The MacA and MacB communicate by the Timbuktu package, which
implements
the TCP/IP protocols over an Ethernet dedicated line.
- switch on the MacB computer (energize the VME crate).
- run Timbuktu on the MacB: you can select it in the Apple (top-left)
menu.
- open the connection selecting the
CONTROL
button after you
entered the TCP/IP number (129.57.188.20
).
- give the username
H2OCEBAF
and the password
WATER
.
The window that appears contains a `carbon copy' of the MacA screen:
the user in counting room can control all the operations of MacA by
using the MacB mouse and keyboard as if the MacB user were using the
MacA; BE CAREFUL: the performance are reduced due to the
connection: do not be in a hurry! wait for the screen to refresh after any
action.
Startup of the system
The system startup depends on the history of the shutdown
After a normal shutdown procedure
- go to the right side of the Right rack and push the green start
button
of the right box; you will hear a beeping sound;
- go to the front of the Left rack and push the green button near the
emergency one;
- wait for the MacA to be operative;
- click on the H2O CEBAF97 icon (the slow control program for the
water target at CEBAF), which is almost in the center of the
MacA desktop. The LabView application will run automatically
and the slow control main window will be shown;
- push the arrow button on the top-left side of this window in order to
start the program.
The first time the program is ran, after MacA power on, it initializes the
electronic modules. Every time the program is ran, it asks for a new
initialization of the electronic modules. Answer YES, a small window
with 4 buttons appears. Press all the buttons in order from top to bottom.
When the initialization is done the main window come back.
After an emergency shutdown procedure
- Recognize the cause of the emergency, and fix it.
- Turn the emergency button in order to restore its operation. The
system will be energized and a beeping sound will be heard.
- Push the green button near the emergency red button. The beeping
sound will be turned off.
System shutdown
The normal procedure
- go to the experimental hall;
- close the program on the MacA; shutdown the MacA;
- move to the right side of the Right rack and push the red
stop button of the right box (the left box is not used).
The emergency procedure
This procedure is really an EMERGENCY procedure; do not use it for normal
operation.
Push the emergengy red button in front of the Left rack if you are in
HAll A, the one in the counting room on the top of the computer monitor.
In order to restart the system follow the startup procedure described above.
Waterfall activating
- check that the slow control program is running on MacA and
optionally on MacB;
- check if the pump is on, and in case turn it on
from the switch in front of Right Rack or from the switch in front of
"Pump Command Module" in Counting Room
- increase the pump speed to about 4/5 of the scale
by moving the slide (pump speed) in the main window of MacA or
MacB (right side);
- wait about 20 minutes to stabilize the water flow, then adjust the
pump speed to the wanted working value
.
Empty target procedure
Put the slide in the right of the main window to the lower position, then
switch off the pump; this can be done from the switch which is in front of
Right Rack or from the switch which is in front panel of the "pump Command
Module",
in Counting Room (just over the MacB monitor); in both cases,
the reactivating of the pump needs to be done from the same switch used
to deactivate it (see "Waterfall Activating Procedure").
Target movement procedure
- press the motor button on the bottom of the main window of the
computer; a motor control window will appear
- select the type of movement (up/down, rotation);
- select the target or the angle; the move button should be activated if
everything is O.K.;
- press the Move button; a warning message remind you to be sure of
your action. Be sure that the beam is off!!
- during the movement watch the two displays showing the current
position and the one to be reached.
- when the final point is reached, a message will inform you of the
success of the operation. Otherwise a warning message will tell you
about the faulty status.
(Eventual) gas flow operating procedures
The gas filling of the target cell, when empty, has to be done in Hall A:
on the
right rack there is a self-eplaining panel. The washing
button has to be
pushed (for about 15 minutes) to open the inlet and outlet valves in order
to allow the
gas to replace the air which fills the target cell.
In order to refill the cell, in the case of very small leakages
(remember that the
pressure must be about 1.02 Atm),
push the pressure bottom part of the main window. The pressure control
window will appear. The operation is self-explaining. Just remember that
the gas flows only if you keep the button pressed.
Calibration of the movement system
For `normal' operation the user does not need any additional information about
the movement system.
The following calibration tables (target and angle positions versus encoder
values) are useful for checking purpose; notice that if the target changes,
the calibration has to be performed again. The following values are OK
for the waterfall, carbon, berillium, iron target which was
installed in July '96.
The corrispondence between the targets and the encoders is as follow:
Target |
Encoder Position |
Water |
17301 |
Up Survey |
38528 |
Carbon |
42949 |
Berillium |
47371 |
Iron |
51792 |
Low Survey |
56213 |
The vertical scale factor is: 1 encoder unit = 5.74 $\mu m$
The angular correspondence is:
Angle |
Encoder Position |
0 degree |
27361 |
30.533 degree |
26479 |
that means: 1 encoder unit = 0.035 degree
When a new calibration is required
only an expert person must perform the following steps:
- go to the Hall A;
- run the calibration program ("setlimits.vi") on MacA in order to
allow the computer to get
the limits of the movement ranges (up/down and rotation); this procedure takes
several minutes;
- rotate the target to two angles and take note of the encoder values
- write these value on a scrapbook
- select sequentially the 6 targets (waterfall and the solid ones)
and take note of the encoders values;
- write them in the table ....
- note ALL the previous values on the logbook.
- move
Next: 5.3.3.18 Troubleshooting
Previous: 5.3.3.14 Safety Assessments
10/13/1997