- Participants: Wilson Miller, Xiangdong Wei, Kevin Wei, Michael Lowry, Alexandre Deur, Andy Sandorfi, Xiaochao Zheng.
- JLab's picture of the L-tube test cell filled with two pellets: SPF L-tube tests Sept 18.pdf
- the jlab setup needed weights to keep the L-tube in place
due to pressure. This should not be necessary if we use plastic tubing
(and cutting a groove around the tube for the O-ring to sit in).
- two pellets were lost/broken during the fill process. In
order to remove static electricity from the pellets for them to drop
into the L-tube, the pellets were washed in ethanol and then pumped in
an oil bath at 200C for 18 hours. Then the 4He permeation at dry-ice
temperature was measured again but it appeared the permeation has
changed w.r.t. before the ethanol wash (page 2 of slides, lower right,
black data points.)
- Wilson sent an updated image of 3 plastic beads (2mm dia), with improved resolution: MR image 9-10-2015.png
- Alexandre will move the complete support structure from JLab to UVa.
- Alexandre will move one pressure-reducing tube to JLab for measuring the pressure drop.
|
- Participants: Wilson Miller, Xiangdong Wei, Kevin Wei, Alexandre Deur, Andy Sandorfi, Xiaochao Zheng.
- Andy's summary on thermal tests: Summary JLab Thermal Tests.pdf The essence is the following:
- we can easily go from 200 C directly to Dry Ice at -78 C in seconds without any problems or signs of damage to the glass.
- the temperature of He gas inside the L-tube follows the
change in bath temperature, but a little more slowly - time constant of
28 sec.
- there is negligible mixing of the hot gas at the bottom
of the L-tube with the cold gas above it. As a result a pressure
differential will be created as we cool the pellet, so we need to make
sure we don't buckle the shell as we cool the L-tube.
- The buckling pressure for the thin shells at 200 C is 3.8
atm, but goes to 21 atm at Dry Ice. We need to keep the pressure
differential across the pellet wall below about 1/2 the buckling
pressure to have some safety margin. (The thicker walled shell could go
higher.)
- With the thin shells we should start off trying to fill
to 4 atm at 200 C. That would certainly be safe. I will do a proper
calc taking into account.
the time constants for cooling the gas.
- Andy's calculation on pressure due to cooling: Differential pressure from cooling.pdf
- I've calculated the time evolution of the pressures
within the pellets when going from 200 C to either dry ice (top panels
in the attached plot) or to LN2 (bottom panels). The theoretical
buckling pressure limits are shown as the blue (14 micron wall) and red
(26 micron) curves. The 26 micron wall pellets are never a problem - we
should start with these.
- However, at 8 atm (left panels) the calculated pressures
(red points) come too near the 14 micron limits (blue curves). With the
thinner pellets weshould limit the initial pressure to about 5 atm.
- Design for the multi-stage filling tube:
- Xiaochao's initial sketch: cell_2015_3.png
- Andy measured the volume of one of the L-tubes, see L-tube volume.pdf
- the 5.5 cc includes 0.5cc in the thick walled bottom 10
cm length and 5 cc in the upper part that ends in the valve seat). This
is V1 in the sketch.
- The cell part of V2 is 9.0-5.5=3.5cc.
- measured
I.D.: chem-thread is 0.307in or 7.80mm; valve
is
0.215in or 5.46mm on both sides. We want to max the fill pressure (as
long as it is below 4-5 atm). So want to minimize the fill-valve
portion of V2 and maximize V3. Also,Wilson mentioned the O-ring might
pop off in the bottom valve of the original sketch.
- So here is a modified sketch: cell_2015_3c.png. Assuming the "minimized" length of V2 is 2cm, then V2=3.5cc+(2cm)*pi*(5.46mm/2 to
7.8mm/2)^2 = 3.5cc+(0.47cc to 0.96cc) = 3.97cc to 4.46cc; V3=(20cm)pi(5.46mm/2)^2 = 4.68 cc.
- Estimate of fill pressure using the calculated volume: If
we fill V3 with 8atm, the filling pressure would be
8atm*(V3)/(V1+V2+V3)=8atm*(4.68)/(5.5+3.97+4.68) to(4.68)/(5.5+4.46+4.68) = (0.33 to 0.32) * 8atm. People
seem to be happy with this ratio.
- Xiangdong's coil support design:
- drawing: MRI Coil Form Assembly-Half View.pdf
- video: Explosion1_1.wmv
- Wilson
updated from the imaging side: gas self-diffusion
may be the ultimate limit on imaging resolution. This effect should be
temperature-dependent but we are not sure how much it will change.
Wilson is trying to
find a way to deal with this, aiming for 0.5mm in all directions.
- We talked about safty: need a pexi-glass shield, face-shields, ear plugs, gloves.
- We talked about the test sequence:
- setup the L-tube, the filling valves, the manifold from the polarizer;
- put the L-tube in the oil bath, pump out the system;
- let the polarized 3He in from polarizer, fill the pellet in multiple stages;
- close the valve plunge the L-tube into dry-ice, then
either measure the cell, or open the valve again, use manifold to pump
out gas outside the pellet, for imaging the pellet.
- disconnect from the manifold, walk down the hall for imaging;
- If we measure the filled cell, the polarization outside
the pellet is supposed to be similar to the polarization in the
polarizer.
- We will start with 2 pellets, so if one breaks, the other still works.
- We will repeat this process at liquid nitrogen
temperature. There has been some evidence that T1 would drop in the
temperatue range of 70-100K due to increased dwell time of 3He on the
wall surface and thus a longer exposure to paramagnetic impurities, see
this figure from Price & Haeberli NIM-A349(94)321. We need to find at what temperature this effect becomes significant.
|
- Participants: Wilson Miller, Xiangdong Wei, Andy Sandorfi, Xiaochao Zheng.
- General updates since last meeting:
- The two broken cells had been repaired and shipped to Andy.
- Andy has ordered some glass tubing from Mike to check the thermal shock.
- Andy showed a presentation: 3He-pellet-conf-Aug21-topics.pdf
- page 2: calculated buckle pressure is 4.8 atm for the thinnest pellet. Would like to keep pressure < 2atm for safety.
- page 5: summary of all permeation measurement for 4He.
All three batches of pellets gave similar results. The measurement was
done at 5 temperatures: dry ice, 0C, room, 100C and 200C, per pellet,
and using 2-3 different methods.
- page 6: 3He and 4He have almost identical permeation. snapshot: 10sec at 200C, 2hrs at dry ice, and 317 years at LN2
- We discussed the test procedure (page 1 of Andy's presentation). A few discussions:
- At some point we discussed pumping 3He out of the test
cell and image the pellet only. The current plan is to start with
imaging the filled cell, and pellet-only image will be measured only if
necessary.
- Andy suggested we start with 2atm test cell first.
- We need to fill the cell in <2atm steps. The plan is
to design a small two-valve system and control the filling by volume
ratio.
- Question: how much density is needed for a good image? Wilson: not sure, but last test was done with 3-4atm at room temp.
- Question: why is T1 measured so short, 45min? Calculation showed tens of hours (see 4/24/2015 minutes).
Wilson: could be O2 (manifold of the polarizer is leaky); in his
experience with this polarizer (bags, cells, etc.) never saw T1 longer
than 1 hour. Impression is it is limited by wall relaxation.
- To do list:
- Jlab will continue working on flushing the cell and design of the thermal system;
- Wilson will continue working on improving the image resolution;
- Design the two-valve system -- Xiaochao
- info: polarizer cell is about 4"-dia sphere with volume 340mL. Our test cell is about 5cc.
- figure out details of the dry ice, LN2 etc at physics -- Xiaochao
- Gordon has a 10L thermos we might be able to borrow.
Demo lab has two, 5L and 10L, but occasionally need both for the
classrooms; LN2 from the physics dewar is about $1/liter; Need a PATEO#
if we take any.
- we don't have dry ice in the department. Need to order if needed. Demo lab sometimes order 30-70 lbs that cost ~$40.
- next meeting will be Friday Sept. 4th.
|
- Participants: Wilson Miller, Xiangdong Wei, Kevin Wei, Andy Sandorfi.
- Wilson
has made a first image of the L-shape tube. 3-4atm room-temperature
polarized 3He, T1~45min. See meeting minutes by Andy here: 3He fusion conf July 31 summar.pdf
|
- Participants: Wilson Miller, Xiangdong Wei, Kevin Wei, Andy Sandorfi, Xiaochao Zheng, (Alexandre joined near the end).
- Andy visited Wilson's lab on June 17. Here are pictures of the coil (with cables) and the cell with two glass beads in.
- Andy, Xiangdong and Kevin have been doing the permeation test in the past three weeks.
- Repeated the measurement for different pressure and
temperatures. The result did not change much compare to the 6/5
meeting. Concluded the present test setup limits reaching higher
precisions so will stop here. The permeation time is similar to what
were reported in the literature.
- Prepared individual pellets: put them in to test vials
(one pellet per vial) and flushed with 4He gas. Wilson pointed out O2
could be a problem here due to the high depolarization effect -- 0.1
amg O2 would cause a 20-s T1. Measured the O2 content with an RGA
(residue gas analyzer) of the 4He gas used and found ~0.1% O2 for the
"ultra pure" and even higher for the "ultra ultra pure" helium gas. The
O2 in 3He gas used for the hyperpolarizer is not a problem since it
reacts out with the alkali vapor right away, but for our test we need
to leave the He-flushed test tubes pumping a long time (probably
overnight)
- one pellet (out of 12) exploded/shattered. Don't know how
it happened. But fortunately no damage was done and no safety report
was required.
- Todo:
- continue the flushing practice.
- calculate the permeation time of O2 to determine the minimum pumping time for the actual test.
- Wilson has been tuning the RF coils
- The current RF coil works should work with a linear polarization adaptor box
("LP" box) connecting to the MRI setup. Has both a LP and a circular
polarization (CP) box but for some reason the LP box does not work
correctly. (Here the LP and CP refers to the x and y component of the
excitation B1 field).
- When trying to fill the glass cells with pellets,
found it was difficult to turn the valve on two of the three cells, and
eventually these two cracked. Here is a picture of the cracked valve. Could it be that the glass-blowing process altered the valve
somewhat?
- Filled the unbroken glass cell with two beads.
- Todo:
- will make the LP box work
- first will try measuring the thermal 3He polarization
in the cell, by filling the cell with 1atm of 3He and 1atm O2. Here the
O2 is to decrease T1 so the thermal polarization can be reached as soon
as the cell is placed in the MRI setup. Hopefully these can be
done within a week. Then will try the hyper-polarized 3He.
- Also discussed the test: we should have the glass cells
tested (empty) at UVa first, then flush/fill with 4He at JLab, then
back to UVa for the final test.
- Other todos:
- Xiaochao will talk to Mike Souza about the broken cells. Will see if Mike or Al can fix them (if we purchase new valves).
|
- Participants: Andy Sandorfi, Xiaochao Zheng; (Wilson is out of
town, Xiangdong is at Disneyland, Alexandre's location is unknown).
- Andy's update on the permeation test: permeation-June5update.pdf. We discussed the following:
- page 1 shows the relevant formula. D: diameter of pellet; w: wall thickness of pellet; kappa: permeation constant.
- On page 2, the left graph is for a 2atm filled pellet,
the right is for am empty pellet. There is a linear term in pressure
due to off-gasing of the pellet and the chamber.
- For the filled pellet test (p2 left), it is hard to get a good fit with the (exponential+linear) function.
- page 3: fit to various time ranges shows tau=(540\pm 30)sec at 297K, consistent with literature
- To dos:
- measure chamber off-gasing without the pellet. This
will help to quantify how much off-gasing is from the pellet (and how
it will eventually affect our polarized pellet)
- will repeat the test for other pellets and at different
temperatures (we should know the temperature dependence of the
off-gasing too)
- Detailed information on the glass material from Mike Souza:
- Here is the glass handbook from Corning.
"Pyrex" is a trademark from Corning when they introduced their
borosilicate glass in 1915. However, they also used "pyrex" for glasses
made from other materials.
- The specific glass used for our test cells is Corning 7740
- Information relevant to the thermal and mechanical properties are on page 828 (Table 4), 834 and 838, and are summarized below:
- In Table 4, the properties of Corning 7740 are the same
as (or at least very similar to) the table summarized below under
"5/29" minutes.
- On page 834, the equation to calculate thermal stress
differ from Xiaochao's simple model by 2(1-nu) where nu=0.20-0.23 for
borosilicate is called Poisson's ratio. This causes a factor 1.6 higher
thermal shock than Xiaochao's calculation. However, this is almost
irrelevant because both the simple and the handbook equations depend on
the ultimate strength of the glass, which in reality varies from sample
to sample. I would suggest that we adhere to the max thermal shock of
160K (3.2mm thick) to 90K (12.7mm thick) shown in the handbook.
- More details of the thermal shock test are given: The
max thermal shock of 160K (3.2mm) to 130K (6.4mm) to 90K (12.7mm
thick) were obtained by plunging a 15x15cm^2 plate from a hot
temperature into a cold reservior. All surfaces are cooled uniformity.
This means the stress is between the center and the surface which for
the 160K result is effectively across a 1.6mm thickness. Our test cell
wall is probably close to 2-3mm thick which means we might want to
limit the shock to 100K.
- Xiaochao will bring some pyrex test tubes to JLab to check thermal shock
- Xiaochao will calculate the time constant of the glass
test cell, to determine how long we need to keep it in boiling
water/ice water etc.
|