Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name V. Sulkosky
Log entry time 19:32:29 on April 21, 2009
Entry number 267253
This entry is a followup to: 267243
keyword=Re: EDTM time offset, L-HRS
A measurement on the scope of the time between the frontend output
of the EDTM (which goes to discriminators) and the backend output,
which goes to BigBite.
The backend output is earlier by 118 ns for the master (S2 EDTM).
The internal delay of the slave EDTM for S1 is 190 ns.
I measured the additional delay due to using the back output of the slave
for BigBite to be 24 ns, this includes the 6 ns cable between the master
output and slave input.
Total delay for BigBite EDTM signal on L-HRS is
16 ns cable
8 ns 754 module
5 ns cable
260 ns RG-58 cable #5
A copy of this log entry has been emailed to: moffit,miham