    User name A. Camsonne

    Log entry time 21:06:22 on June 15, 2009

    Entry number 281622

    Followups:

    keyword=Compton electron detector work

    Today, we traced down the problem of the left detector ( odd strip numbers ), to a filtering ferrite which seemed to have failed ( maybe because of the power supply short ). It was by passed and the detector seemed to work though only tested with noise. We double checked that laser status was recorded correctly. Though I still need to figure out the bad data flagged as laser off in runs from previous night.

    I also put the analog board on plane A. The signals goes to a 428 linear fan which are sent to the scope on adaqs1 and to 10 times amplifier. The signal of cable 19 is split on another fan out and sent upstairs on cable 19 and to the flash ADC on the second channel, it would be good to have configuration to read out this channel or both photon and analog signal. Cable 14 has the other strip and is on the scope and sent upstairs through 10 times amplifier.