Main INDEX
Monthly INDEX
PREV
NEXT
User name R. Michaels
Log entry time 17:17:14 on November 1,2001
Entry number 75575
This entry is a followup to: 75491
keyword=EDTM system fixup
Bodo and I realized that the reason the scaler measurement of EDT
didn't make sense was that we are pulsing 8 channels, but only 6 are
used by scintillators. The extra 2 can overlap with trigger, leading to
a falsely "negative" deadtime. The best solution would be to disable the
extra 2 channels, or to form the "or" of only the 6 that are used. We didn't
have time to do this during this access, so we simply send all 8 channels
through the trigger now. The EDT measured in scalers will be approx
correct now, except for a small correction due to the 2 channels having
no scintillator signals. Meanwhile the event analysis should work, and
we can now compare with scalers.