Main INDEX
Monthly INDEX
PREV
NEXT
User name P. Monaghan
Log entry time 23:47:47 on January 17, 2005
Entry number 139540
This entry is a followup to: 139521
keyword=auxiliary plane discriminator threshold change
During the access earlier this evening we changed the auxiliary plane
discriminator threshold from -40mV to -100mV (see halog entry 139521). By
plotting the adc spectrum for an auxiliary plane bar (#20) with and
without a cut on the tdc spectrum, we have verified that the increase in
threshold only affects (and cuts out) the pedestal. This is illustrated
in the plot shown below; the black line is without the tdc cut and the
green line is after the tdc cut is applied.
Figure 1 - ADC plot for auxiliary plane bar #20