We had an opportunistic access to the left HRS during the accelerator
hall A access. Looking at the A1 PMT 10 (11th channel) which is dead in
the histograms (both ADC and TDC), I found there was no significant
signal on the oscilloscope. Turning up the HV in small increments to
2900 did not improve things much. [The signal became visible but had a
small amplitude. The current drawn went to a "normal" value of 500 at
2900 V.]
The tube appears bad. Turned the HV back to previous, and left the hall.
When we have a chance we should replace the PMT (monday or tuesday this
week, as soon as a new tube is prepared).