    User name R. Michaels

    Log entry time 00:04:17 on July21,2005

    Entry number 148806

    Followups:

    keyword=profile scanner checkout

    Since the profile scanner scintillator was allegedly fixed, we
    will do a few short runs to check versus VDCs.

    Run 2187:
    On both arms, X = 15, Y = 34 cm
    Turn on HV for scint. at -1800 V
    See T5 = 16.8 Hz and T6 = 15 Hz

    Run 2188:
    Remains in same X,Y location. ADC spectra looked low on R-arm.
    Increase HV on scint: L-arm to 1850 and R-arm to 1900.
    Notice the T5,T6 rates don't change much.

    Run 2189
    X = 18, Y = 30

    Run 2190
    X = 15, Y = 28

    Fig 1 shows the ADC spectra for T5 & T6 for run 2188.
    Not much amplitude in the R-arm. But then we got something
    for run 2189 -- fig 2. There's an X-Y spot too which I'll show
    shortly.

    0:03 End of profile scanner checkout.



    FIGURE 1

    FIGURE 2