    User name R.Lindgren

    Log entry time 13:56:23 on November13,2005

    Entry number 160032

    Followups:

    keyword=Magnets down

    1:55 PM Q1 right won't ramp up. Tech discovered helium flow meter on Q1 right is frozen (No pun intended). Tech found that Q1 temperature is high and vacuum is bad. While trouble shooting, the flow meter reset tself to 0, then the helium flow valve opened to Q1. This put a load on the helium system and as a result we lost all the HRS magnets and the septa due to helium level faults. Tech zeroed the magnets to prevent a fast dump of the magnets. Right now Q1 is being evacuated and the helium cooling system is stablizing. Temperature of Q1 is coming down. Cryo person on call was also called in by Crew chief.