    User name spiegel

    Log entry time 08:59:05 on November14,2005

    Entry number 160085

    This entry is a followup to: 160032

    keyword=Tech's

    Verified Q1 ramp problem. Reset magnet. No change in ramp problem. Discovered that Q1 magnet vacuum was bad. and magnet temp. was 72 deg.K. Realised Q1 magnet Helium level was false. Meter readback was 100%. Reset meter indication went to zero %. Ramped down other magnets to minimize load on helium system. Began vacuum pump down of Q1 magnet. Restarted cooldown. Q1 He liquid level meter indicated a level again as it filled back up. However, level meter was replaced for reliablity. Reset magnets. And began ramp up after cooldown recovery.