User name R. Michaels
Log entry time 03:28:37 on November22,2005
Entry number 161008
This entry is a followup to: 161006
Followups:
keyword=another rate calibation, saturation observed
Actually run 3137 does NOT make sense !!
Run 3138: Try again at lower currents
Event 21K see 1 uA
Event 62K ask for 0.5 uA
Event 84K see 0.5 uA
Request 0.2 uA
Event 177K see 0.2 uA
Event 218K ask for 2.0 uA
Event ~250K have 2.0 uA
We can see the "saturation" in rate online. Rates are semi-linear
up to ~0.5 uA then saturates at 2 MHz. I think it may be due to
the discriminator output which is 100 nsec --> max rate 10 MHz.
Presumably we have significant electronic deadtime. FWIW, the
scalers are rated 200 MHz. We could probably do a little better
next time with narrow (10 nsec) pulses.