• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name Target Group

    Log entry time 15:41:02 on December 24, 2008

    Entry number 254550

    This entry is a followup to: 254466

    keyword=NMR Water Calibration

    A water calibration of 2000 NMR sweeps are performed from Dec 23 to Dec 24. A relative precision of 1.2% is achieved on the NMR signal.

    Sweep Settings:

    NMR Settings for Upstream Coil
    --
    Target Type:	Water
    --
    Holding Field (G):	18.018353
    Theta (degree):	90.000000
    Phi (degree):	0.042174
    High Field (G):	25.000000
    Sweep Speed (G/s):	1.200000
    Sit Time (s):	5.000000
    RF Frequency (kHz):	90.700000
    RF Power (Vrms):	0.500000
    --
    Start Time:	2008/12/23 20:07:56
    Finish Time:	2008/12/24 14:06:13
    Wait Time (m):	0.000000
    Number of Sweeps:	2000
    --
    Lockin Sensitivity:	12
    Time Constant:	4
    Phase (degree):	0.000000
    Sample Rate:	10
    TC Filter:	4
    Wide Reserve:	2
    Close Reserve:	2
    

    ===============

    NMR Settings for Downstream Coil -- Target Type: Water -- Holding Field (G): 18.018353 Theta (degree): 90.000000 Phi (degree): 0.042174 High Field (G): 25.000000 Sweep Speed (G/s): 1.200000 Sit Time (s): 5.000000 RF Frequency (kHz): 90.700000 RF Power (Vrms): 0.500000 -- Start Time: 2008/12/23 20:07:56 Finish Time: 2008/12/24 14:06:13 Wait Time (m): 0.000000 Number of Sweeps: 2000 -- Lockin Sensitivity: 12 Time Constant: 4 Phase (degree): 0.000000 Sample Rate: 10 TC Filter: 4 Wide Reserve: 2 Close Reserve: 2






    NMR Sweep



    Analysis