PMT 17 is pretty noisy. (Been true for a while. We didn't want to swap it out since it will require a longish down to flush the entire Cerenkov tank.)
HVs for PMTs 2*,4,8,9,10,12* were tweaked. (*) changed by >50V Previous HVs recorded here (run 1810): http://www.jlab.org/~adaq/halog/html/0902_archive/090225223803.html New HVs can be found here (run 1841): http://www.jlab.org/~adaq/halog/html/0902_archive/090226170318.html Run 1829: - HVs for 1--10 on, pulser on - LED #11, #15, #20 plugged in - LED drive pulse ampl. set to provide 1--2 p.e. pulse on the scope for PMT 10 - Pedestal run - T8 trigger dominated (others prescaled away) = peds: 483, 407, 398, 479, 543, 499, 603, 527, 460, 465 Run 1830: - HVs for 1--10 on, pulser on - LED #11, #15, #20 plugged in = aligning 1--8 to pedestal + 30 = aligning 9--10 to pedestal + 60 (since they are not split) = adjusted PMTs 2,3,4,8 = adjusted PMTs 9, 10 Run 1831: - HVs for 1--10 on, pulser on - LED #11, #15, #20 plugged in = adjusted PMT 4 Run 1832: - HVs for 1--10 on, pulser on - LED #11, #15, #20 plugged in = adjusted PMT 4 Run 1833: - HVs for 1--10 on, pulser on - LED #11, #15, #20 plugged in - raised LED drive voltage a bit to enhance 2 p.e. peaks = looks OK Run 1834: - HVs for 11--20 on, pulser on - LED #1, #5, #10 plugged in - adjusted LED drive voltage for this LED group - Pedestal run - T8 trigger dominated (others prescaled away) = peds: 303, 425, 515, 578, 579, 443, 410, 542, 329, 412 Run 1835: - HVs for 11--20 on, pulser on - LED #1, #5, #10 plugged in = adjusted 12 Run 1836: - HVs for 11--20 on, pulser on - LED #1, #5, #10 plugged in - LED drive was too large, lowered = 12 good Run 1837: - HVs for 11--20 on, pulser on - LED #1, #5, #10 plugged in - LED drive was still too large for some of the PMTs, lowered it again = Gains look pretty good. = PMT 17,18 have much noisier pedestals than the rest, will investigate that now.