Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name R. Michaels
Log entry time 17:01:17 on November 8,2009
Entry number 298368
Followups:
keyword=scaler gating issues
Diancheng, Ramesh, Bob
Took access and found the GMN2 signal unplugged and dangling in
front of the NIM/ECL that sent it elsewhere. We plugged it back in.
This means the R-arm data we were taking was bad because we were
integrating over the helicity transition period (it has transients).
I have to leave now, while Ramesh and Diancheng sort out the
polarity issues. It seems there are some flipped cables, etc.
Hopefully they will post a follow-up.
Just to remind about scalers: The 4th input is a VETO. This means
if the signal is logical-true it won't count. (If you have no signal
there it will count forever.) From this you can tell what polarity
is needed.
We've agreed that a future diagnostic will be to plot the measured
integration time minus expected value. If 0 or 1 count, it's fine.
If much more (or less ?) it's a problem.