User name Saha
Log entry time 12:17:29 on October 27,1998
Entry number 3835
This entry is a followup to: 3822
keyword=Hall A BCM linearity calibration
We finally did the Hall A BCM linearity and calibration procedures.
Linearity Test: We cycled the beam through 36, 19.6, 10.6, 5.3, 1.91, 1.1, .46, .2 microA in steps of ~90 seconds each. The data is logged under Run #2070.
Standard Calibration: We cycled the beam through 35.5 microA for 6 cycles about 60 seconds each. The data is logged under Run #2072.