Feb 22, 2021

From Hall A Wiki
Jump to: navigation, search

Back to Main Page >> Moller Page >> Moller Weekly Meeting Index

previous meeting << >> following meeting


Moller DocDB

Original Moller Polarimeter Page

Moller ELog


Agenda:

Analysis progress:

  • Kerr results: Don Jones and Kate Evans [1]
  • Proposed beam pipe size changes Bill Henry [2]

Connection details:

Meeting URL https://bluejeans.com/388466836

Meeting ID 388 466 836

US toll free phone 1.888.240.2560

Dial in, type in the meeting ID and then press ##


Minutes

Discussed option of resizing beam pipe near Moller detector and adding corrector coils. Bill Henry presented studies he did with G3 simulation that show a moderate increase in backgrounds mostly due to reduced shielding. Nothing absolutely conclusive, but it appears that we would be largely insensitive to the additional backgrounds since they are low in energy.

Kate Evans and Don Jones discussed Kerr measurements and showed that they can't seem to achieve saturation. Even "null" results with a silvered mirror showed field dependence. Dave Gaskell suggests terminating the signal from the photodiode with 50 ohms to ensure highest bandwidth and lowest frequency dependence.