Gemeel122 log 101 200

From Hall A Wiki
Jump to: navigation, search

2nd set (Run # 101-200)

RUN # Source/Cosmic/Pedestal Source Position Time HV # of events notes
101 S Low X, Low Y 3700 Bad reading in Y
102 S High X, High Y 3700 No Data recorded - ROC Froze
103 S "" "" No Data recorded - ROC Froze
104 S "" "" Reset ROC
105 S High X, Low Y 3700 Bad reading in Y
106 S High X, Mid Y "" Y Reading Fine
107 S High X, Low-Mid y "" Y Reading Bad
108 P Changed Y Card - Card was not connected
109 P
110 S High X, Low-Mid Y Y Reading Bad
111 ABORTED
112 S Center Old readout Card - It is bad.
113 S Center ROC FROZE
114 Checked CH1 readout - Nothing
115 Checked CH0
116 S Mid X, High Y Y reading Fine
117 P GEM 3<----- PLOTS
118 S Center 3700
119 S Low X, Low Y "" Y reading bad
120 S High X, High Y "" Y Reading fine
121 P Redid Ped Run
122 S High X, Low Y 3700 Y reading Bad
123 P Switched Boards on GEM
124 S High X, Low Y 3700 Y reading Bad
125 Changed Amplifier Output to Slot 2 and messed with DC Voltage
126 P
127 S High X, Low Y 3700 Y reading Bad
128 P GEM 4 <----- PLOTS
129 C 3600 BAD
130 S Center 3700
131 S Low X, Low Y 3700 Y Reading Bad
132 S High X, Mid Y 3700 Y reading good
133 S High X, Mid-Low Y "" Y reading Bad
134 S "" 3750 Y reading Bad
135 S High X, High Y 3700 Y reading good
136 S High X, High Y 3700 Y reading Good, Higher up on Y
162 P GEM 4. Amplification board 1
163 P GEM 4. Amplification board 2, no difference w/ board 2
164 P GEM 4. Amp board 1. changed T.gem.x1.detmap & T.gem.y1.detmap->slot_lo to 1 in database
165 S 3590 bottom->GEM 4 (source). top->GEM2. 2.73mA
166 S 3600 GEM 4
167 S GEM4
168 P cram.gem.crl T1 70->60
169 P cramsTprog T1 60->80
170 P T1 80->70

T

171 P no input connected. Test to see what problem is
172 S Low X, Low Y 3700
173 aborted
174 sS Low X, Low Y 3700 changed to channel 2
175 P replaced Y card and conector
176 P changed CRAM card
177 S Mid X, Mid Y 3700
178 P GEM 2
179 C 3600
180 changed T.gem.X1.detmap & T.gem.Y1.detmap ->slot_lo to 1
181 S changed T.gem.X1.detmap & T.gem.Y1.detmap ->slot_lo to 2 (changed back)
182 S
183 changed T1 to 50 in CRL code for 551b
184 P (kaptons unpluged)
185 P
186 S 3700 Changed T1 from 50 to 40 in CRL code
187 S 3700 CRL code still changed to 40. Data is going to neg(-) input in 550
188 P Changed polarity of trigger. Hold -> no difference
189 S 3700
190 S 3700 Changed CRL T1->70
191 3700 Changed CRL T1->60
192
193 Changed CRL T1-> 50. aborted
194 gate generator to make trigger hold. T1->50
195 S 3700 T1 50 -> 30
196 S 3700 T1 30 -> 20
197 3750 T1 20 -> 70
198 Clear coming from gate generator
199 P
200 S 3750