Transversity Target Calibration List
This is a collection of he3 target calibration measurement for experiment transversity
Please add an entry for each furture calibration. Please include
- Time stamp as second order title
- the data directory
- Result if there is any
- measurement condition
- Link to Halog Entry
- Recommend to include the measurement time line into Halog entry, NOT this WIKI
Contents
- 1 Cell History
- 2 EPR Calibration
- 2.1 Brady: Transverse Pumping, L-HRS: Negative
- 2.2 Brady: Vertical Pumping, L-HRS: Negative
- 2.3 Brady: Vertical Pumping, L-HRS: Positive
- 2.4 Brady: Transverse Pumping, L-HRS: Positive
- 2.5 Maureen: Vertical Pumping, L-HRS: Positive
- 2.6 Maureen: Transverse Pumping, L-HRS: Positive
- 2.7 Maureen: Transverse Pumping, L-HRS: Negative
- 2.8 Maureen: Vertical Pumping, L-HRS: Negative
- 2.9 Astral: Vertical Pumping, L-HRS: Negative
- 2.10 Astral: Transverse Pumping, L-HRS: Negative
- 3 Water Calibration
- 4 Internal Temp. Calibration
- 5 NMR Cross Calibration
- 6 Spin Up Test
- 7 AFP Loss Test
Cell History
ASTRAL: Oct 20 - Nov 25, 2008
- Cell Name: ASTRAL
- He3 Density (amg): 8.08
- N2 Density (amg): 0.11
- Vpc (cc): 164.92
- Vtt (cc): 6.77
- Vtc (cc): 79.47
- NMR Lock-in gain: x10
- Data : click me
MAUREEN: Nov 26 - Dec 24 2008
- Cell Name: MAUREEN
- He3 Density (amg): 7.23 (7.52)
- N2 Density (amg): 0.102 (0.106)
- Vpc (cc): 180.75
- Vtt (cc): 4.16
- Vtc (cc): 89.05
- NMR Lock-in gain: x5
- Data : click me
BRADY: Jan 7 - Feb 4 2009
- Cell Name: BRADY
- He3 Density (amg): 7.87
- N2 Density (amg): 0.11
- Vpc (cc): 169.27
- Vtt (cc): 5.98
- Vtc (cc): 74.57
- NMR Lock-in gain: x5
- Data : click me
EPR Calibration
Brady: Transverse Pumping, L-HRS: Negative
Feb 3, 2009
- Cell: Brady
- Holding Field: Transverse - (LHRS:- RHRS:0 )
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25x2 + FAP 25)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090203_0944
- Halog Entry: Click me
- Calibration Constant: 2.77%/mV
Feb 3, 2009
- Cell: Brady
- Holding Field: Transverse - (LHRS:- RHRS:0 )
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25x2 + FAP 25)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090203_0937
- Halog Entry: Click me
- Calibration Constant: 2.74%/mV
Jan 28, 2009
- Cell: Brady
- Holding Field: Transverse + (LHRS:- RHRS:0 )
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090128_1853
- Halog Entry: Click me
- Calibration Constant: 2.81%/mV
Brady: Vertical Pumping, L-HRS: Negative
Feb 5, 2009
- Cell: Brady
- Holding Field: Vertical + (LHRS:- RHRS:0 )
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25x2 + FAP 25)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090205_0738
- Halog Entry: Click me
- Calibration Constant: 4.47%/mV
Feb 5, 2009
- Cell: Brady
- Holding Field: Vertical - (LHRS:- RHRS:0 )
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25x2 + FAP 25)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090205_0730
- Halog Entry: Click me
- Calibration Constant: 4.59%/mV
Feb 3, 2009
- Cell: Brady
- Holding Field: Vertical + (LHRS:- RHRS:0 )
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx2 + FAP 25W)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090203_1257
- Halog Entry: Click me
- Calibration Constant: 4.55%/mV
Feb 3, 2009
- Cell: Brady
- Holding Field: Vertical + (LHRS:- RHRS:0 )
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx2 + FAP 25W)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090203_1246
- Halog Entry: Click me
- Calibration Constant: 4.44%/mV
Jan 28, 2009
- Cell: Brady
- Holding Field: Vertical + (LHRS:- RHRS:0 )
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090128_1655
- Halog Entry: Click me
- Calibration Constant: 4.67%/mV
Jan 23, 2009
- Cell: Brady
- Holding Field: Vertical + (LHRS:- RHRS:0 )
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090123_1514
- Halog Entry: Click me
- Calibration Constant: 4.73%/mV
Jan 23, 2009
- Cell: Brady
- Holding Field: Vertical + (LHRS:- RHRS:0 )
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090123_1503
- Halog Entry: Click me
- Calibration Constant: 4.58%/mV
Brady: Vertical Pumping, L-HRS: Positive
Jan 20, 2009
- Cell: Brady
- Holding Field: Vertical + (LHRS:+ RHRS:0 )
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090120_1141
- Halog Entry: Click me
- Calibration Constant: 4.45%/mV
Jan 18, 2009
- Cell: Brady
- Holding Field: Vertical + (LHRS:+ RHRS:0 )
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090118_1744
- Halog Entry: Click me
- Calibration Constant: 4.53%/mV
Brady: Transverse Pumping, L-HRS: Positive
Jan 12, 2009
- Cell: Brady
- Holding Field: Transverse - (LHRS:+ RHRS:0 )
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090112_1754
- Halog Entry: Click me
- Calibration Constant: 2.83%/mV
Jan 11, 2009
- Cell: Brady
- Holding Field: Transverse - (LHRS:+ RHRS:0 )
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20090111_1507
- Halog Entry: Click me
- Calibration Constant: 2.79%/mV
Maureen: Vertical Pumping, L-HRS: Positive
Dec 22, 2008
- Cell: Maureen
- Holding Field: Vertical - (LHRS:+ RHRS:0 )
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081222_1203
- Halog Entry: Click me
- Calibration Constant: 5.16%/mV
Dec 22, 2008
- Cell: Maureen
- Holding Field: Vertical - (LHRS:+ RHRS:0 )
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081222_1117
- Halog Entry: Click me
- Calibration Constant: 5.03%/mV
Dec 16, 2008
- Cell: Maureen
- Holding Field: Vertical - (LHRS:+ RHRS:0 )
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081216_1002
- Halog Entry: Click me
- Calibration Constant: 5.10%/mV
Dec 13, 2008
- Cell: Maureen
- Holding Field: Vertical - (LHRS: Positive polarity)
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081213_1807
- Halog Entry: Click me
- Calibration Constant: 5.07%/mV
Maureen: Transverse Pumping, L-HRS: Positive
Dec 13, 2008
- Cell: Maureen
- Holding Field: Transverse - (LHRS: Positive polarity)
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081213_0859
- Halog Entry: Click me
- Calibration Constant: 2.98%/mV
Dec 5, 2008
- Cell: Maureen
- Holding Field: Transverse - (LHRS: Positive polarity)
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081208_1509
- Halog Entry: Click me
- Calibration Constant: 2.99%/mV
Maureen: Transverse Pumping, L-HRS: Negative
Dec 5, 2008
- Cell: Maureen
- Holding Field: Transverse - (LHRS: Negative polarity)
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081205_0903
- Halog Entry: Click me
- Calibration Constant: 2.92%/mV
Dec 2, 2008
- Cell: Maureen
- Holding Field: Transverse -
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081202_0833
- Halog Entry: Click me
- Calibration Constant: 2.92%/mV
Nov 28, 2008
- Cell: Maureen
- Holding Field: Transverse -
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081128_1520
- Halog Entry: Click me
- Calibration Constant: 2.92%/mV
Maureen: Vertical Pumping, L-HRS: Negative
Nov 28, 2008
- Cell: Maureen
- Holding Field: Vertical +
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081128_1008
- Halog Entry: Click me
- Calibration Constant: 5.08%/mV
Nov 26, 2008
- Cell: Maureen
- Holding Field: Vertical +
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081126_1028
- Halog Entry: Click me
- Calibration Constant: 5.24%/mV
Nov 26, 2008
- Cell: Maureen
- Holding Field: Vertical +
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081126_1016
- Halog Entry: Click me
- Calibration Constant: 5.23%/mV
Astral: Vertical Pumping, L-HRS: Negative
Nov 25, 2008
- Cell: Astral
- Holding Field: Vertical +
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081125_0846
- Halog Entry: Click me
- Calibration Constant: 2.48%/mV
Nov 20, 2008
- Cell: Astral
- Holding Field: Vertical +
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081120_1145
- Halog Entry: Click me
- Calibration Constant: 2.43%/mV
Nov 18, 2008
- Cell: Astral
- Holding Field: Vertical +
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081118_1859
- Halog Entry: Click me
- Calibration Constant: 2.40%/mV
Astral: Transverse Pumping, L-HRS: Negative
Nov 18, 2008
- Cell: Astral
- Holding Field: Transverse - (Correction Coil ON: Small 0.18A)
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081118_1744
- Halog Entry: Click me
- Calibration Constant: 1.44%/mV
Nov 18, 2008
- Cell: Astral
- Holding Field: Transverse - (Correction Coil ON: Small 0.18A)
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081118_1730
- Halog Entry: Click me
- Calibration Constant: 1.45%/mV
Nov 18, 2008
- Cell: Astral
- Holding Field: Transverse -
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081118_1711
- Halog Entry: Click me
- Calibration Constant: 1.44%/mV
Nov 18, 2008
- Cell: Astral
- Holding Field: Transverse -
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081118_1650
- Halog Entry: Click me
- Calibration Constant: 1.44%/mV
Nov 15, 2008
- Cell: Astral
- Holding Field: Transverse -
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081115_1427
- Halog Entry: Click me
- Calibration Constant: 1.48%/mV
Nov 14, 2008
- Cell: Astral
- Holding Field: Transverse -
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081114_1704
- Halog Entry: Click me
- Calibration Constant: 1.46%/mV
Nov 13, 2008
- Cell: Astral
- Holding Field: Transverse -
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx2+FAP 25Wx1)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081113_1121
- Halog Entry: [Click me
- Calibration Constant: 1.48%/mV
Nov 11, 2008
- Cell: Astral
- Holding Field: Transverse -
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 100W (FAP 20Wx5)
- Data Directory:
C:\HappyValentine\EPR\Data\EPR_AFP_20081111_0916
- Halog Entry: Click me
- Calibration Constant: 1.45%/mV
Water Calibration
Oct.4, 2008
- data files:
NMR_Water_20081004_221633 NMR_Water_20081004_221914 NMR_Water_20081004_222533
See halog: http://www.jlab.org/~adaq/halog/html/0810_archive/081004223759.html
- survey photos:
/work/halla/transversity/disk2/Target/WaterNMR/10.05
Oct.11, 2008
- data files:
NMR_Water_20081011_233531 NMR_Water_20081012_174209
- survey photos:
/work/halla/transversity/disk2/Target/WaterNMR/10.12
Dec.23, 2008
- data files:
NMR_Water_20081223_195848 NMR_Water_20081223_200756
see halog: http://www.jlab.org/~adaq/halog/html/0812_archive/081224154102.html
- survey photos:
/work/halla/transversity/disk2/Target/WaterNMR/12.24 Xmas Water
Mar.23, 2009
- data files:
Group 1 (3699 sweeps): NMR Data: C:\HappyValentine\NMR\Data\NMR_Water_20090320_205008 Field Data: C:\HappyValentine\VI\Field Rotation\Calibration\Field reading 03-20-2009 20-45-33.txt Group 2 (3699 sweeps): NMR Data: C:\HappyValentine\NMR\Data\NMR_Water_20090322_111205 Field Data: C:\HappyValentine\VI\Field Rotation\Calibration\Field reading 03-22-2009 11-10-59.txt
see halog: http://www.jlab.org/~adaq/halog/html/0903_archive/090323121956.html
- survey photos:
/work/halla/transversity/disk2/Target/WaterNMR/other_pic.tar.gz
Internal Temp. Calibration
Feb.3,2009
- Cell Brady
- Transverse -
http://www.jlab.org/~adaq/halog/html/0902_archive/090203114627.html
Jan 28,2009
- Cell Brady
- Vertical +
http://www.jlab.org/~adaq/halog/html/0901_archive/090128175037.html
Jan 20,2009
- Cell Brady
- Vertical +
http://www.jlab.org/~adaq/halog/html/0901_archive/090120130316.html
Jan 17,2009
- Cell Brady
- Transverse -
- Old method
http://www.jlab.org/~adaq/halog/html/0901_archive/090117132035.html
Jan 12,2009
- Cell Brady
- Transverse -
- UVA method from Al
Jan 11, 2009
- Cell: Brady
http://www.jlab.org/~adaq/halog/html/0901_archive/090111155650.html
Dec 22, 2008
- Cell: Maureen
- Oven Temperature: 230C
- Laser Power: 75 W(Comet 25Wx3) , OFF before calibration
- Calibrated Field Direction: Vertical +
- Calibrated NMR RF Field: Longitudinal
- Calibrated NMR Pumping Chamber Pickup Coil: Transverse
- DATA ::: C:\HappyValentine\NMR\Data\NMR_He_20081222_221340
- Time Line:
1:23 Laser Off 1:40 Test Freq Sweep to get field strenth (since hall truck ramp is open). NMR_Freq_He_20081222_134031 Peak at 84.25kHz 1:43 Freq Sweep NMR Data ID : NMR_Freq_He_20081222_134320 ScreenShotNMR_FrequencySweep_v6.vi_22-12_13-44-24.html 1:45 Frield Rotation to Long. + 1:48 Target -> Pickup Coil 1:56 Field Sweep NMR @ Pick up Data ID : NMR_He_20081222_135619 NMRField_22-12_13-56-54.html 1:56 Move Target Back to He3 1:58 Frield Rotation to Vert. + 2:09 Freq Sweep Data ID : NMR_Freq_He_20081222_140851 ScreenShotNMR_FrequencySweep_v6.vi_22-12_14-11-27.html 2:10 Laser Back
Dec 13, 2008
- Cell: Maureen
http://www.jlab.org/~adaq/halog/html/0812_archive/081214075556.html
Nov 20, 2008
- Cell: Astral
http://www.jlab.org/~adaq/halog/html/0811_archive/081121010827.html
Nov 14, 2008
- Cell: Astral
http://www.jlab.org/~adaq/halog/html/0811_archive/081114174438.html
Nov 5, 2008
- Cell: Astral
http://www.jlab.org/~adaq/halog/html/0811_archive/081105232737.html
(note::::: This test was performed with the FAP lasers. The first two ( i.e. on Nov. 20 and on Nov. 14 ) tests were performed with COMET lasers)
http://www.jlab.org/~adaq/halog/html/0901_archive/090112193858.html
NMR Cross Calibration
Jan 28, 2009
- Cell: Brady
- Holding Field: Vertical +
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Time Line
18:01 NMR Freq Sweep @ Trans Data ID : NMR_Freq_He_20090128_180134 18:01 Field -> Long. +; Target -> Pick up 18:07 NMR Field Sweep @ Pickup Data ID : NMR_He_20090128_180717 18:07 Field -> Trans -; Target -> he3 18:13 NMR Freq Sweep @ Trans Data ID : NMR_Freq_He_20090128_181329
- Data Dir
C:\HappyValentine\NMR\Data\NMR_Freq_He_20090128_180134 C:\HappyValentine\NMR\Data\NMR_He_20090128_180717 C:\HappyValentine\NMR\Data\NMR_Freq_He_20090128_181329
- Halog Entry: Click me
Jan 20, 2009
- Cell: Brady
- Holding Field: Vertical +
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20090120_115442 C:\HappyValentine\NMR\Data\NMR_He_200900120_120441 C:\HappyValentine\NMR\Data\NMR_Freq_He_20090120_121308
- Halog Entry: Click me
Jan 18, 2009
- Cell: Brady
- Holding Field: Transverse -
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20090118_104320 C:\HappyValentine\NMR\Data\NMR_He_20090118_104948 C:\HappyValentine\NMR\Data\NMR_Freq_He_20090118_105608
- Halog Entry: Click me
Jan 11, 2009
- Cell: Brady
- Holding Field: Transverse -
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20090111_170004 C:\HappyValentine\NMR\Data\NMR_He_20090111_170721 C:\HappyValentine\NMR\Data\NMR_Freq_He_20090111_171422
- Halog Entry: Click me
Dec 13, 2008
- Cell: Maureen
- Holding Field: Transverse -
- Oven Temperature: 230C
- details in the halog entry
- Data Directory:
C:\Happy Valentine\NMR\Data\NMR_Freq_He_20081213_100659 C:\Happy Valentine\NMR\Data\NMR_He_20081213_102104 C:\Happy Valentine\NMR\Data\NMR_Freq_He_20081213_103239
- Halog Entry: Click me
Nov 28, 2008
- Cell: Maureen
- Holding Field: Vertical +
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081128_103303 C:\HappyValentine\NMR\Data\NMR_He_20081128_104433 C:\HappyValentine\NMR\Data\NMR_Freq_He_20081128_105403
- Halog Entry: Click me
Nov 25, 2008
- Cell: Astral
- Holding Field: Vertical +
- Pumping Direction: +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081125_090237 C:\HappyValentine\NMR\Data\NMR_He_20081125_091713 C:\HappyValentine\NMR\Data\NMR_Freq_He_20081125_093210
- Halog Entry: Click me
Nov 20, 2008
- Cell: Astral
- Holding Field: Vertical +
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081120_122106 C:\HappyValentine\NMR\Data\NMR_He_20081125_124603 C:\HappyValentine\NMR\Data\NMR_Freq_He_20081125_131112
- Halog Entry: Click me
Nov 14, 2008
- Cell: Astral
- Holding Field: Transverse -
- Pumping Direction: -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081114_1254xx C:\HappyValentine\NMR\Data\NMR_He_20081115_0114xx C:\HappyValentine\NMR\Data\NMR_Freq_He_20081115_0127xx
- Halog Entry: Click me
Spin Up Test
Nov 14, 2008
- Cell: Astral
- Holding Field: Transverse -
- Pumping Direction: -
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081114_034713
- Halog Entry: Click me
- Spin Up Time: 4.65 +- 0.07 hours
- Max NMR Value: 48.20 +- 0.21 mV
Nov 14, 2008
- Cell: Astral
- Holding Field: Transverse -
- Pumping Direction: +
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081114_210850
- Halog Entry: Click me
- Spin Up Time: 5.03 +- 0.28 hours
- Max NMR Value: 48.65 +- 0.55 mV
Jan 11, 2009
- Cell: Brady
- Holding Field: Transverse -
- Pumping Direction: -
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
C:\HappyValentine\NMR\Data\NMR_He_20090110_161254
- Halog Entry: Click me
- Spin Up Time: 3.82 +- 0.11 hour
- Max NMR Value: 17.32 +- 0.08 mV = 65.7%
AFP Loss Test
Nov 26, 2008
- Cell: Maureen
- Holding Field: Vertical +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
Auto spin flip, sequence 259, flip count# 4-13
- Halog Entry: Click me
- AFP Loss per Flip: 0.50%
Nov 25, 2008
- Cell: Astral
- Holding Field: Vertical +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
Auto spin flip, sequence 253, flip count# 411-420
- Halog Entry: Click me
- AFP Loss per Flip: 0.65%
Nov 20, 2008
- Cell: Astral
- Holding Field: Vertical +
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Transverse
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
Auto spin flip, sequence 253, flip count# 72-81
- Halog Entry: Click me
- AFP Loss per Flip: 0.50%
Nov 18, 2008
- Cell: Astral
- Holding Field: Transverse -
- NMR RF Coil: Longitudinal
- NMR Pumping Chamber Pickup Coil: Vertical
- Oven Temperature: 230C
- Laser Power: 75W (COMET 25Wx3)
- Data Directory:
Auto spin flip, sequence 248, flip count# 9-18
- Halog Entry: Click me
- AFP Loss per Flip: 0.54%