Difference between revisions of "Thermal Equilibrium Measurement Procedure"

From Hall A Wiki
Jump to: navigation, search
(Prepare Fridge:)
 
(15 intermediate revisions by 2 users not shown)
Line 1: Line 1:
;This procedure was created for GEN, and must be updated to reflect changes in the target for g2p.
 
  
TO aid in the TE calculation, use the "te_getter.pl" script in the g2p_te folder on poltarg2p2.
+
===Software===
  
Changes for g2p: Set the run valve to a reasonable value, close to the setpoint chosen by the PID. 65% to 90% is a bit too much for this fridge.
+
To aid in the TE calculation, use the "te_getter.pl" script in the g2p_te folder on poltarg2p2. This script reads the last 'x' data entries before unixtime 'y' (default is current time).  It produces a datafile for these points and asks if you'd like to use them in a TE measurement.  Then it performs the TE calculation and returns the results.
  
 
===Assumptions:===
 
===Assumptions:===
Line 9: Line 8:
 
*Magnet is at full field, persistent mode, leads are ramped down
 
*Magnet is at full field, persistent mode, leads are ramped down
 
*Nose is full, run valve in manual mode
 
*Nose is full, run valve in manual mode
===Prepare Fridge:===
+
===Prepare Frig:===
*Stop RB3, RB2, and RB1 (wait 2 mintues between each), if necessary
+
*Stop RB3, RB2, and RB1 (wait 2 minutes between each), if necessary
 +
*Close roughing and bypass valve
 
*Put Run Valve, EV91120, in Manual Mode
 
*Put Run Valve, EV91120, in Manual Mode
*Establish a STEADY Nose level somewhere between 65% and 90%
+
*Establish a STEADY Nose level somewhere between 50-65%
*Only make slight changes to Run Valve for rest of TE (+/- 0.05 each 10 min)
+
*Try 0.17 on the run valve.
 +
*If needed only make slight changes to Run Valve for rest of TE (+/- 0.05 each 10 min)
 +
 
 
===Take a Baseline:===
 
===Take a Baseline:===
 
*Hit "Unlock Magnet Controls" button
 
*Hit "Unlock Magnet Controls" button
*Type the full-field current in the box labeled "Setpoint" (eg: 77.085)
+
*Type the full-field current in the box labeled "Setpoint" (eg: 60.882)
*Type 0.50 into "Setrate" box, if necessary
+
*Type 3.00 into "Setrate" box, if necessary
 
*Hit "To Setpoint" Button
 
*Hit "To Setpoint" Button
 
*Wait for leads to reach full current (approx 1 min)
 
*Wait for leads to reach full current (approx 1 min)
Line 24: Line 26:
 
*Hit "Heater On" button
 
*Hit "Heater On" button
 
*Wait for the 30 second timeout to expire
 
*Wait for the 30 second timeout to expire
*Type the baseline current into the "Setpoint" box (eg: 74.600)
+
*Type the baseline current into the "Setpoint" box (eg: 59.882)
 
*Hit "To Setpoint" to start the magnet sweeping
 
*Hit "To Setpoint" to start the magnet sweeping
 
*Wait for magnet to reach baseline current (approx 6 min)
 
*Wait for magnet to reach baseline current (approx 6 min)
Line 31: Line 33:
 
*Hit the Baseline button
 
*Hit the Baseline button
 
*Select "Create New Baseline", dialog box should then disappear
 
*Select "Create New Baseline", dialog box should then disappear
*Change sweeps to 5000 if necessary
+
*Change sweeps to 2000 if necessary
 
*Double check that all NMR settings are where you want them
 
*Double check that all NMR settings are where you want them
 
*Hit "One Point" button to take a single nmr measurement
 
*Hit "One Point" button to take a single nmr measurement
Line 39: Line 41:
 
*Document the details of the baseline in the logbook:
 
*Document the details of the baseline in the logbook:
 
**Date Time, #Sweeps, MagCurrent, Top/Bottom, Gain, RFFreq, RFMod
 
**Date Time, #Sweeps, MagCurrent, Top/Bottom, Gain, RFFreq, RFMod
*Type full-field current into "Setpoint" box (eg: 77.085)
+
*Type full-field current into "Setpoint" box (eg: 60.882)
*Type 0.50 into "Setrate" box, if necessary
+
*Type 3.00 into "Setrate" box, if necessary
 
*Hit "To Setpoint" button
 
*Hit "To Setpoint" button
*Wait for magnet to reach full current (approx 6 min)
+
*Wait for magnet to reach full current
 
*Hit "Hold" button
 
*Hit "Hold" button
 
*Hit "Heater Off" button
 
*Hit "Heater Off" button
Line 51: Line 53:
 
*Make sure ladder is in desired target position
 
*Make sure ladder is in desired target position
 
*Make sure NMR is on desired channel (AND in agreement with target position!)
 
*Make sure NMR is on desired channel (AND in agreement with target position!)
*Set sweeps to 5000, if necessary
+
*Make sure you have a STEADY Nose level with roughing and bypass valve closed
 +
*Use ~0.17 on the run valve (watch the nose temps.)
 +
*Set sweeps to 2000, if necessary
 
*Hit "Take Data" button
 
*Hit "Take Data" button
 
*Wait for timer to count down
 
*Wait for timer to count down
 
*Write the following in the logbook for the next 10 measurements:
 
*Write the following in the logbook for the next 10 measurements:
 
**Time, NMR Area, 4He Press, 4He Temp, 3He Press, 3He Temp, Nose Level
 
**Time, NMR Area, 4He Press, 4He Temp, 3He Press, 3He Temp, Nose Level
 +
*Watch data until it flattens out, if in good warm (~1.5K) semi-steady state for a while then it should take less than 30 minutes.
 +
*calculate the TE calibration constant using poltar@poltarg2p2$ /home/poltar/g2p_te/te_getter.pl 15
 +
*Type Y to run te_calc on this data
 +
*Move the output file te_result.txt to te_date_top/bot.txt
 +
*Put new Cal. Const. in under Sweeps in PDP and move on to prepare frig for polarization

Latest revision as of 07:55, 17 April 2012

Software

To aid in the TE calculation, use the "te_getter.pl" script in the g2p_te folder on poltarg2p2. This script reads the last 'x' data entries before unixtime 'y' (default is current time). It produces a datafile for these points and asks if you'd like to use them in a TE measurement. Then it performs the TE calculation and returns the results.

Assumptions:

  • NMR System is tuned
  • Magnet is at full field, persistent mode, leads are ramped down
  • Nose is full, run valve in manual mode

Prepare Frig:

  • Stop RB3, RB2, and RB1 (wait 2 minutes between each), if necessary
  • Close roughing and bypass valve
  • Put Run Valve, EV91120, in Manual Mode
  • Establish a STEADY Nose level somewhere between 50-65%
  • Try 0.17 on the run valve.
  • If needed only make slight changes to Run Valve for rest of TE (+/- 0.05 each 10 min)

Take a Baseline:

  • Hit "Unlock Magnet Controls" button
  • Type the full-field current in the box labeled "Setpoint" (eg: 60.882)
  • Type 3.00 into "Setrate" box, if necessary
  • Hit "To Setpoint" Button
  • Wait for leads to reach full current (approx 1 min)
  • Hit "Hold"
  • Check that leads current and magnet current are equal
  • Hit "Heater On" button
  • Wait for the 30 second timeout to expire
  • Type the baseline current into the "Setpoint" box (eg: 59.882)
  • Hit "To Setpoint" to start the magnet sweeping
  • Wait for magnet to reach baseline current (approx 6 min)
  • Hit "Hold" button
  • Put NMR into pause mode if necessary
  • Hit the Baseline button
  • Select "Create New Baseline", dialog box should then disappear
  • Change sweeps to 2000 if necessary
  • Double check that all NMR settings are where you want them
  • Hit "One Point" button to take a single nmr measurement
  • Wait for timer to count down
  • Hit Baseline button
  • Select the baseline you just took from the list of timestamps
  • Document the details of the baseline in the logbook:
    • Date Time, #Sweeps, MagCurrent, Top/Bottom, Gain, RFFreq, RFMod
  • Type full-field current into "Setpoint" box (eg: 60.882)
  • Type 3.00 into "Setrate" box, if necessary
  • Hit "To Setpoint" button
  • Wait for magnet to reach full current
  • Hit "Hold" button
  • Hit "Heater Off" button
  • Wait for 30 second timeout to expire
  • Hit "To Zero" to ramp the leads down
  • Hit "Lock Magnet Controls" button

Take TE measurements:

  • Make sure ladder is in desired target position
  • Make sure NMR is on desired channel (AND in agreement with target position!)
  • Make sure you have a STEADY Nose level with roughing and bypass valve closed
  • Use ~0.17 on the run valve (watch the nose temps.)
  • Set sweeps to 2000, if necessary
  • Hit "Take Data" button
  • Wait for timer to count down
  • Write the following in the logbook for the next 10 measurements:
    • Time, NMR Area, 4He Press, 4He Temp, 3He Press, 3He Temp, Nose Level
  • Watch data until it flattens out, if in good warm (~1.5K) semi-steady state for a while then it should take less than 30 minutes.
  • calculate the TE calibration constant using poltar@poltarg2p2$ /home/poltar/g2p_te/te_getter.pl 15
  • Type Y to run te_calc on this data
  • Move the output file te_result.txt to te_date_top/bot.txt
  • Put new Cal. Const. in under Sweeps in PDP and move on to prepare frig for polarization