4/5/2012    pedestal      3761/22788/40837
 
 
3/28/2012    75nA      3437/22470/40486
 
 
3/16/2012    50nA      3254/22238/40388
 
 
3/13/2012    25nA      3248/22338/40368
 
 
3/10/2012    pedestal  3053/22133/40296
 
  
 
===Analysis and scripts===
 
===Analysis and scripts===
 
+
you can find the script in:
 
/adaqfs/home/adaq/pzhu/bcm/tungsten
 
/adaqfs/home/adaq/pzhu/bcm/tungsten
  
 
===Tungsten Calorimeter Calibration Results  ===
 
===Tungsten Calorimeter Calibration Results  ===
 +
calibration constant can find in: [http://hallaweb.jlab.org/experiment/g2p/collaborators/pzhu/wiki/pyDB/bcm%20calibration.ods here]
 +
 +
Note: SIS3801 have different constant with SIS3800
 +
 +
~3.17 3rm downstream scaler abnormal
 +
3.18~4.2 LHRS upstream scaler noisy
 +
4.2~4.9 RHRS SIS3801 scaler not working
 +
4.9 changed RHRS scaler channel for bcm
 +
4.19~ calibration constant changed
 +
Near 5.12 3rm sis3801 not working?
 +
5.13~5.14 downstream broken
 +
 +
in constant area:
 +
blue available slope
 +
green available pedestal
 +
black not available
 +
 +
"in avail run area:
 +
  black means not available
 +
same color  in neighbour cell means same calibration constant"
  
  happex I=slope*(A-ped)
+
how to use calibration constant to calulate charge:
        C=slope*875us*(ΔA-ped*ΔEntry))
+
  happex C=slope*875us*(ΔA-ped*ΔEntry))
  scaler C=slope*(ΔA-ped*ΔEntry)
+
  scaler SIS3800 & SIS3801 C=slope*(ΔA-ped*ΔEntry)
  
calibration constant can find in: [https://docs.google.com/spreadsheet/ccc?key=0AgIgVNT0l_ttdHNRZjQ4S0k3bzJfLUoyYTRJRFpVYXc#gid=0 here]
+
how to use calibration constant to calulate current:
 +
happex: I=slope*(rate-ped)*875/1041.65
 +
sis3800: I=slope*(rate-ped*clockrate)
 +
sis3801: I=slope*(rate-ped*103.7k*971.65us)/1041.65us
  
 
==New Hall A 2 channel BCM receiver 1H00==
 
==New Hall A 2 channel BCM receiver 1H00==

Latest revision as of 04:37, 19 November 2012

Tungsten calorimeter

Introduction

Analysis and scripts

you can find the script in: /adaqfs/home/adaq/pzhu/bcm/tungsten

Tungsten Calorimeter Calibration Results

calibration constant can find in: here

Note: SIS3801 have different constant with SIS3800

~3.17	3rm downstream scaler abnormal
3.18~4.2	LHRS upstream scaler noisy
4.2~4.9	RHRS SIS3801 scaler not working
4.9	changed RHRS scaler channel for bcm
4.19~	calibration constant changed
Near 5.12	3rm sis3801 not working?
5.13~5.14	downstream broken
in constant area:	
blue	available slope
green	available pedestal
black	not available
"in avail run area:
 black means not available
same color  in neighbour cell means same calibration constant"	

how to use calibration constant to calulate charge:

happex	C=slope*875us*(ΔA-ped*ΔEntry))
scaler	SIS3800 & SIS3801 C=slope*(ΔA-ped*ΔEntry)

how to use calibration constant to calulate current:

happex:  I=slope*(rate-ped)*875/1041.65
sis3800: I=slope*(rate-ped*clockrate)
sis3801: I=slope*(rate-ped*103.7k*971.65us)/1041.65us