Difference between revisions of "Solid calorimeter readout"
From Hall A Wiki
(→signal estimation for preshower) |
(→MAPMT) |
||
Line 32: | Line 32: | ||
= MAPMT = | = MAPMT = | ||
− | + | * file location | |
+ | ** general at https://www.dropbox.com/sh/365le2o93vpn14e/2at980iq-y/readout/MAPMT | ||
+ | ** LHCb related at https://www.dropbox.com/sh/365le2o93vpn14e/g_v9IHvaLF/other_calorimeter/LHCb | ||
− | MAPMT | + | * Our Study |
− | + | ** W&M study [https://www.dropbox.com/sh/365le2o93vpn14e/5W0btJ3NzY/readout/MAPMT/study/MAPMT_gain_match_report.pdf] | |
+ | ** SDU study [https://www.dropbox.com/sh/365le2o93vpn14e/OOquAlIy75/readout/MAPMT/HAMAMATSU%20MAPMT%20summary.pdf] [https://www.dropbox.com/sh/365le2o93vpn14e/fUpSqobRFj/readout/MAPMT/MAPMT%20spec.xlsx] | ||
− | + | * LHCb | |
− | + | ** LHCb structure before ECAL is 1.5cm SPD + 2.5X Lead + 1.5cm PS, MAPMT H7546 with special front end electronic for readout, 10bit ADC for PS, 1bit for SPD | |
− | + | ** LHCb Hamamatsu R5900-00-M64 (12 stages) test in 2005 [https://www.dropbox.com/sh/365le2o93vpn14e/AmWFi3VZDc/other_calorimeter/LHCb/1-s2.0-S0168900204020145-main.pdf], 2000 [http://lhcb-calo.web.cern.ch/lhcb-calo/internal/TDR/notes/031/bt.ps] | |
− | + | ** LHCb Hamamatsu R7600-00-M64MOD (8 stages) mass characterization in 2007 [https://www.dropbox.com/sh/365le2o93vpn14e/gPv3AVW0R1/other_calorimeter/LHCb/1-s2.0-S0168900206021474-main.pdf] | |
− | + | ** H7546 as a package may have involved from R5900-00-M64 to R7600-00-M64MOD inside | |
− | + | ||
− | + | ||
− | + | ||
− | + | * Hamamatsu info | |
− | http:// | + | ** MAPMT assembly from hamamatsu http://www.hamamatsu.com/jp/en/product/category/3100/3002/index.html |
− | + | ** H7546 is a package name and its current spec sheet mentions R7600-M64 PMT inside[https://www.dropbox.com/sh/365le2o93vpn14e/qNmztVpj9k/readout/MAPMT/Hamamatsu/H7546A_H7546B_TPMH1240E12.pdf] | |
+ | ** The new MA-PMT, H12700 [https://www.dropbox.com/sh/365le2o93vpn14e/N38qy6-LBx/readout/MAPMT/Hamamatsu/New%20MA-PMT%20from%20Hamamatsu.pdf], is still under development. We hope to be able to release it by the end of this year | ||
+ | ** H8500's uniformity map in Figure 3 on page 2 of its datasheet (http://www.hamamatsu.com/resources/pdf/etd/H8500_H10966_TPMH1327E02.pdf). By standard spec, the non-uniformity has to be less than 1:3. It reflects variation in anode sensitivity, which is gain x photocathode sensitivity. The 1:3 means that we record the relative output of all anodes: the lowest output must be larger than 1/3 of the highest one. | ||
− | + | * other | |
− | + | ** CLAS12 RICH H8500C test, including crossstalk result (Hoek_H8500_Characterization.pdf) | |
− | + | ** CLAS12 RICH has some H7546 test done also (Montgomery_NPESeminar_150911.pdf) | |
− | + | ** CLAS12 RICH frontend (130626_RICH_Tech_Rev_FrontEnd.pdf) | |
− | + | ** CLAS12 RICH DAQ (HallB_RICH_DAQ_June2013.pdf) | |
− | + | ** CLAS12 RICH review report (Rich_Tech_Review_report.pdf) | |
− | + | ** SBS coordinate director, similar to SoLID EC SPD (CDet-talk-SBS-MeetingJune2013-Sarty.pptx) | |
− | CLAS12 RICH has some H7546 test done also | + | |
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + |
Revision as of 14:01, 5 November 2013
file dir
on dropbox
https://www.dropbox.com/sh/365le2o93vpn14e/yM3akCdLSK/readout
signal estimation for preshower
- assumption
- 22 photoelectrons per MIP from LHCb test in 2005 [1], 105 photoelectrons per MIP from LHCb test around 2000 [2]. We take the later result as 22. Our 2cm preshower scintilator is 33% thicker than 1.5cm in LHCb PS, so we assume 30 photoelectrons for SoLID PS MIP
- in SoLID preshower, one MIP is 6MeV, max signal is 200MeV, (refer to the plot in SoLID pCDR). If we plan to cover 6 - 240MeV which is 1 - 40 MIP, then photoelectrons from PS range is 30-1200
- pulse full width 20ns according to scope picture from Simona's test of H8500C (slide 26 of talk). According to LHCb study, pulse shape has many variation and 25ns only covers about 85% of pluses on average. We assume 30ns.
- Simplify pluse shape as a triangle shape, then pulse height is about 2 times pulse area
- 50 ohm matching
- at 1000V, MAPMT H10966 has gain 3e5 (comparing to MAPMT H8500C has gain 1.5e6)
- both MAPMT H10966 and H8500C has anode uniformity (=gain*photocathode sensitivity) 1:3 in its spec sheet (Simona's test confirms it) and this leads to a factor 9. H7546 has 1:3 according to LHCb tesr, but it has 1:5 according its spec. We assume 1:3
- result without uniformity problem
- pulse height min 30*3e5/30e-9*1.6e-19*50*2=0.0048V
- pulse height max 5600*3e5/30e-9*1.6e-19*50*2=0.192V
- dynamic range 40
- result with 1:3 uniformity problem
- pulse height min 30*3e5/30e-9*1.6e-19*50*2/3=0.0016V
- pulse height max 1200*3e5/30e-9*1.6e-19*50*2*3=0.576V
- dynamic range 360
- conclusion
- pulse height max < flashADC limit 2V
- 12 bit ADC with 4096 channel like the Jlab flasADC can cover the dynamic range required about 400 if a MIP is at channel 10
MAPMT
- file location
- LHCb
- LHCb structure before ECAL is 1.5cm SPD + 2.5X Lead + 1.5cm PS, MAPMT H7546 with special front end electronic for readout, 10bit ADC for PS, 1bit for SPD
- LHCb Hamamatsu R5900-00-M64 (12 stages) test in 2005 [6], 2000 [7]
- LHCb Hamamatsu R7600-00-M64MOD (8 stages) mass characterization in 2007 [8]
- H7546 as a package may have involved from R5900-00-M64 to R7600-00-M64MOD inside
- Hamamatsu info
- MAPMT assembly from hamamatsu http://www.hamamatsu.com/jp/en/product/category/3100/3002/index.html
- H7546 is a package name and its current spec sheet mentions R7600-M64 PMT inside[9]
- The new MA-PMT, H12700 [10], is still under development. We hope to be able to release it by the end of this year
- H8500's uniformity map in Figure 3 on page 2 of its datasheet (http://www.hamamatsu.com/resources/pdf/etd/H8500_H10966_TPMH1327E02.pdf). By standard spec, the non-uniformity has to be less than 1:3. It reflects variation in anode sensitivity, which is gain x photocathode sensitivity. The 1:3 means that we record the relative output of all anodes: the lowest output must be larger than 1/3 of the highest one.
- other
- CLAS12 RICH H8500C test, including crossstalk result (Hoek_H8500_Characterization.pdf)
- CLAS12 RICH has some H7546 test done also (Montgomery_NPESeminar_150911.pdf)
- CLAS12 RICH frontend (130626_RICH_Tech_Rev_FrontEnd.pdf)
- CLAS12 RICH DAQ (HallB_RICH_DAQ_June2013.pdf)
- CLAS12 RICH review report (Rich_Tech_Review_report.pdf)
- SBS coordinate director, similar to SoLID EC SPD (CDet-talk-SBS-MeetingJune2013-Sarty.pptx)