== Move target during production (automatic spin flip) == | == Move target during production (automatic spin flip) == | ||
− | * '''NEVER''' move target during a | + | * '''NEVER''' move target during a flip |
* Move target from production ( He3 ) to other locations | * Move target from production ( He3 ) to other locations |
Revision as of 22:58, 8 November 2008
Move target while spin-up curve test
- NEVER move target during a sweep
- Time for next sweep can be estimated by adding wait time to last sweep time
Move target during production (automatic spin flip)
- NEVER move target during a flip
- Move target from production ( He3 ) to other locations
- If there is no production run, skip to step 4
- Stop current data taking
- Do Flip NOW
- Turn OFF(pause) Auto Flip
- Move target to new position
- Move target to production ( He3 ) from other locations
- Move target to He3 position for production
- Turn ON(pause) Auto Flip
- Do Flip NOW
- Once flip is done, start to take data