Difference between revisions of "Compton Photon Detector Centering Procedure"
From Hall A Wiki
m (Cornejo moved page Spring 2019 Compton Commissioning to Compton Photon Detector Centering Procedure: More generic name) |
(→Compton Photon Detector Centering Procedure) |
||
Line 1: | Line 1: | ||
− | = | + | == Centering Beam on Laser == |
* Before MCC begins the initial process of passing bream through the Chicane, we must: | * Before MCC begins the initial process of passing bream through the Chicane, we must: | ||
*# Check and set HV: | *# Check and set HV: | ||
Line 36: | Line 36: | ||
*#* Ideally we would like the slope between BPM 02A and 02B to be as close to zero as possible | *#* Ideally we would like the slope between BPM 02A and 02B to be as close to zero as possible | ||
+ | == Centering Photon Detector == | ||
* After the beam is centered on the laser, ask MCC to lock the beam in the optimal position found previously and then begin the detector centering procedure | * After the beam is centered on the laser, ask MCC to lock the beam in the optimal position found previously and then begin the detector centering procedure | ||
+ | *# Open up the the Detector Centering strip tools | ||
+ | *## in the compton home directory type: <code>StripTools ~/compton/compton_detector_centering.stp</code> | ||
*# Check and make sure all HV Compton channels [[HV_Control_GUI]] are '''ON''' (should have been done by steps above, so just confirm nothing has tripped off) | *# Check and make sure all HV Compton channels [[HV_Control_GUI]] are '''ON''' (should have been done by steps above, so just confirm nothing has tripped off) | ||
*# Start a new run | *# Start a new run |
Revision as of 12:21, 24 June 2019
Centering Beam on Laser
- Before MCC begins the initial process of passing bream through the Chicane, we must:
- Check and set HV:
- PhDet HV -> OFF
- Side (V) and Top (H) Finger Scintillator -> ON
- Laser table scintillators -> ON
- The following EPICS channels may be useful to monitor the backgrounds
- ComptonUSbg1
- ComptonUSbg2 <-- May not be updating on EPICS/StripTools yet
- ComptonDSbg1
- ComptonDSbg2
- The following EPICS values may also be useful
- hac_bcm_average
- IPM1P02A.XPOS
- IPM1P02A.YPOS
- IPM1P02B.XPOS
- IPM1P02B.YPOS
- Optionally the two finger scintillators may be useful, though it's likely they won't see anything if they are behind the collimator
- ComptonHfingerRate
- ComptonVfingerRate
- It may be useful to take a run during this period. Even though the HV on the detector is off, we may use it to study noise on the system.
- Check and set HV:
- After the beam is stable in the chicane and ready to center beam on laser
- Check and set HV:
- PhDet HV -> ON
- Side (V) and Top (H) Finger Scintillator -> ON
- Start a Compton Run (though, no need to analyze the run right now, this will just help us commission and calibrate the DAQ later)
- With the laser ON and 'not flipping, monitor the following EPICS channels:
- IPM1P02A.YPOS
- IPM1P02B.YPOS
- ComptonCentralRate
- ComptonCentralRateHigh
- COMPTON_PW1PCAV_ca
- As the beam is scanned vertically monitor for spikes in ComptonCentralRate and ComptonCentralRateHigh.
- The goal is to maximize ComptonCentralRate (assuming stable beam and laser power) and record the vertical position
- Ideally we would like the slope between BPM 02A and 02B to be as close to zero as possible
- Check and set HV:
Centering Photon Detector
- After the beam is centered on the laser, ask MCC to lock the beam in the optimal position found previously and then begin the detector centering procedure
- Open up the the Detector Centering strip tools
- in the compton home directory type:
StripTools ~/compton/compton_detector_centering.stp
- in the compton home directory type:
- Check and make sure all HV Compton channels HV_Control_GUI are ON (should have been done by steps above, so just confirm nothing has tripped off)
- Start a new run
- Center the detector vertically (Note: do not lower the table to 0, always keep it above 0 mm)
- Set the Photon Detector table to X=69.00 mm, Y=6.00 mm
- Monitor EPICS channels:
- ComptonHfingerRate <-- looking for a maximum/peak on this when scanning table position
- COM_DETPH_YCPOSai
- ComptonCentralRate <-- possibly should see a minimum on this channel
- ComptonCentralRateHigh <-- possibly should see a minimum on this channel
- Then in 0.5 mm increments scan Y from 6.00 to 16.00 mm (~2-3 seconds at a time is enough)
- Once a rough idea of where the max was, scan a range in as small increments as possible with a +/- 2mm range
- Once you confirm the optimal y-position that gives maximum ComptonHfingerRate maximum (and minimum on ComptonCentralRate and ComptonCentralRate) , then move the table up by 20 mm from this position.
- Open up the the Detector Centering strip tools
- Center the detector horizontally (Note: do not lower the table to 0, always keep it above 0 mm)
- Set the Photon Detector table to X=44.00 mm, Y=<value from step above>
- Monitor EPICS channels:
- ComptonVfingerRate <-- looking for a maximum/peak on this when scanning table position
- COM_DETPH_XCPOSai
- ComptonCentralRate <-- possibly should see a minimum on this channel
- ComptonCentralRateHigh <-- possibly should see a minimum on this channel
- Then in 0.5 mm increments scan Y from 44.00 to 54.00 mm (~2-3 seconds at a time is enough)
- Once a rough idea of where the max was, scan a range in as small increments as possible with a +/- 2mm range
- Once you confirm the optimal y-position that gives maximum ComptonVfingerRate maximum (and minimum on ComptonCentralRate and ComptonCentralRate) , then move the table in +x direction by 20 mm from this position.
- Center the detector horizontally (Note: do not lower the table to 0, always keep it above 0 mm)