# Log on to ''adev@adaq1'' or ''adev@adaq2''.
+
# Start ''jmenu'' on the compton
# ''cd slowc''
+
# Type ''hv'' in the search bar
# ''./hvs BEAMLINE''
+
# Select ''HV Card 3'' - the first channel in this card is the photon detector. Nominal voltage 1750 V.
 +
# For the scintillators on the laser table, select ''HV Card 1'', channels 9, 10 , 11, 12. Nominal HV is +1800 V.
  
 +
The information below is for historical reference.
  
This will start the GUI. Available channels should say "S1, S5, S6, S9, S12". Navigate to the table and select a channel. ''Make sure you know the desired voltage before proceeding.'' Check the box to enable, then click the HV On/Off on the left, if it is not already active.
+
These Photon Detector channels are for PREX/CREX running conditions:
 +
{| class="wikitable"
 +
!
 +
! Slot
 +
! Cha
 +
! HV
 +
|-
 +
| Side (V) Finger
 +
| SL-6
 +
| CH-10
 +
| -950 V
 +
|-
 +
| Top (H) Finger
 +
| SL-6
 +
| CH-11
 +
| -1000 V
 +
|-
 +
| GSO (PhDetector)
 +
| SL-12
 +
| CH-11
 +
| '''See note 1''' <!-- -1700 V -->
 +
|}
 +
'''Note 1:''' For PRex the GSO is in an independent HV supply, set to '''-2340 V'''.
  
Currently, we are working only with '''Slot 12, Channel 11, set to about -1690V'''. (desired voltage as of 2016-11-10)
+
In addition, there are four scintillators in the laser table, two upstream and two downstream of the interaction point. '''Note: The following voltages are all positive.'''
 +
 
 +
{| class="wikitable"
 +
!
 +
! Slot
 +
! Channel
 +
! HV
 +
|-
 +
| DS Top
 +
| SL-9
 +
| CH-0
 +
| +1800 V
 +
|-
 +
| DS Bottom
 +
| SL-9
 +
| CH-1
 +
| +1800 V
 +
|-
 +
| US BR
 +
| SL-9
 +
| CH-2
 +
| +1800 V
 +
|-
 +
| US BL
 +
| SL-9
 +
| CH-3
 +
| +1800 V
 +
|}
 +
 
 +
Return to [[Compton]]
 +
 
 +
'''OLD INSTRUCTIONS'''
 +
 
 +
# Log on to ''adev@adaq1'' or ''adev@adaq2''.
 +
# cd slowc
 +
# ./hvs BEAMLINE
  
 +
Old channel for PbWO4: Slot 12, Channel 11, set to about -1690V. (desired voltage as of 2016-11-10 to 2016-12-31)
  
 
Return to [[Compton]]
 
Return to [[Compton]]

Latest revision as of 07:22, 1 May 2025

This GUI is used to enable/set particular voltage channels remotely.


  1. Start jmenu on the compton
  2. Type hv in the search bar
  3. Select HV Card 3 - the first channel in this card is the photon detector. Nominal voltage 1750 V.
  4. For the scintillators on the laser table, select HV Card 1, channels 9, 10 , 11, 12. Nominal HV is +1800 V.

The information below is for historical reference.

These Photon Detector channels are for PREX/CREX running conditions:

Slot Cha HV
Side (V) Finger SL-6 CH-10 -950 V
Top (H) Finger SL-6 CH-11 -1000 V
GSO (PhDetector) SL-12 CH-11 See note 1

Note 1: For PRex the GSO is in an independent HV supply, set to -2340 V.

In addition, there are four scintillators in the laser table, two upstream and two downstream of the interaction point. Note: The following voltages are all positive.

Slot Channel HV
DS Top SL-9 CH-0 +1800 V
DS Bottom SL-9 CH-1 +1800 V
US BR SL-9 CH-2 +1800 V
US BL SL-9 CH-3 +1800 V

Return to Compton

OLD INSTRUCTIONS

  1. Log on to adev@adaq1 or adev@adaq2.
  2. cd slowc
  3. ./hvs BEAMLINE

Old channel for PbWO4: Slot 12, Channel 11, set to about -1690V. (desired voltage as of 2016-11-10 to 2016-12-31)

Return to Compton