Difference between revisions of "Polhe3 Metal Windows"
From Hall A Wiki
(→Window Pressure Test Setup) |
(→Window Pressure Test Setup) |
||
Line 1: | Line 1: | ||
=== Window Pressure Test Setup === | === Window Pressure Test Setup === | ||
− | * Investigating Larson glass to metal seals as means to couple metal windows to glass target chamber. | + | * Investigating the possibility of using Larson glass to metal seals as means to couple metal windows to glass target chamber. |
− | * Pressure test setup assembled at UVa to test | + | * Pressure test setup was assembled at UVa to test glass to metal seals at typical target cell operating pressures. |
− | * Initial test on a Larson CP-100-T Pyrex glass to Copper metal seal performed on Nov 16 & 17, 2001. Test piece was pressurized to approximately 10-atm (137psig) of argon for 24hrs. The glass to metal seal did not explode. | + | * Initial test on a Larson CP-100-T Pyrex glass to Copper metal seal (1-inch O.D.) performed on Nov 16 & 17, 2001. Test piece was pressurized to approximately 10-atm (137psig) of argon for 24hrs. The glass to metal seal did not explode. Although, 10-atm is low (typical target operating pressure is 12-atm to 13-atm). |
− | + | ||
{| | {| | ||
| [[Image:Ptest1_cp-100-T_16nov2011.jpg|thumb|upright|220px|UVa pressure test setup for target windows. Larson CP-100-T glass to metal seal attached.]] | | [[Image:Ptest1_cp-100-T_16nov2011.jpg|thumb|upright|220px|UVa pressure test setup for target windows. Larson CP-100-T glass to metal seal attached.]] |
Revision as of 20:07, 29 November 2011
Window Pressure Test Setup
- Investigating the possibility of using Larson glass to metal seals as means to couple metal windows to glass target chamber.
- Pressure test setup was assembled at UVa to test glass to metal seals at typical target cell operating pressures.
- Initial test on a Larson CP-100-T Pyrex glass to Copper metal seal (1-inch O.D.) performed on Nov 16 & 17, 2001. Test piece was pressurized to approximately 10-atm (137psig) of argon for 24hrs. The glass to metal seal did not explode. Although, 10-atm is low (typical target operating pressure is 12-atm to 13-atm).
File:Ptest8 cp-100-T 16nov2011.jpg Test pressure gauge P=140psig. |
File:Ptest9 cp-100-T 16nov2011.jpg Regulator gauges: Right to left: Tank pressure P=600psig. Output pressure P=137psig. |