Difference between revisions of "Prepare Frig for Polarization"
From Hall A Wiki
(New page: ;This procedure has been updated to reflect changes in the target for g2p but has not yet been edited by target coordinator. ===Software=== To aid in the TE calculation, use the "te_get...) |
|||
Line 1: | Line 1: | ||
− | ;This procedure | + | ;This procedure has not yet been edited by target coordinator. |
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
===Assumptions:=== | ===Assumptions:=== | ||
*NMR System is tuned | *NMR System is tuned | ||
*Magnet is at full field, persistent mode, leads are ramped down | *Magnet is at full field, persistent mode, leads are ramped down | ||
− | *Nose is | + | *Nose is over 50%, run valve in PID mode |
+ | *Baseline taken and set | ||
+ | *TE has been taken and new calibration constant feed into PDP | ||
+ | *Target in correct position along with NHR Channel setting | ||
+ | *Roughing and Bypass valve open with mechanical pumps on | ||
===Prepare Fridge:=== | ===Prepare Fridge:=== | ||
− | * | + | *Close Roughing valve |
− | * | + | *Turn on Blower 1 and wait 2 minutes |
− | * | + | *Turn on Blower 2 and wait 2 minutes |
− | * | + | *Close Bypass valve |
− | * | + | *Turn on Blower 3 |
− | === | + | *Wait a little while |
− | * | + | ===Begin Polarization:=== |
− | + | *Press "TRANSMIT" on the microwave control panel under Control Unit Status (WARNING: Do not touch the button to the right) | |
− | * | + | *Press "Take Data" on PDP |
− | * | + | *Try to maximize polarization by moving the bellows position |
− | + | *Watch the frequency in the top left monitor | |
− | + | *You can try to start out in the frequency range in the plot of frequency vs Dose since last annel | |
− | + | *Generally, you have to search for the "best" frequency When searching, make changes by 5MHz, 10MHz at the most | |
− | + | *After making a change, wait about 3 NMR updates to see if it helped/hurt | |
− | + | *Beam trips will change pol quickly and obscure the effect of your changes especially over 50nA | |
− | * | + | *If the frequencies display does not work look also at the poltarcama/Jpeg/CamImgs.jpg using Camera Set 0, use F5 to update |
− | * | + | *As radiation damage accumulates, frequency will move if polarizing positively, frequency will move down and if polarizing negatively, frequency will move up |
− | * | + | *Use the logbook − see what frequencies worked for other people |
− | + | *Log all of your best frequencies when you find them in the Target Operator Microwave Frequency Notebook | |
− | + | ||
− | + | ||
− | * | + | |
− | + | ||
− | * | + | |
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | * | + | |
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | * | + | |
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | *Use | + | |
− | + | ||
− | * | + | |
− | + | ||
− | + | ||
− | + | ||
− | + |
Revision as of 03:09, 16 March 2012
- This procedure has not yet been edited by target coordinator.
Assumptions:
- NMR System is tuned
- Magnet is at full field, persistent mode, leads are ramped down
- Nose is over 50%, run valve in PID mode
- Baseline taken and set
- TE has been taken and new calibration constant feed into PDP
- Target in correct position along with NHR Channel setting
- Roughing and Bypass valve open with mechanical pumps on
Prepare Fridge:
- Close Roughing valve
- Turn on Blower 1 and wait 2 minutes
- Turn on Blower 2 and wait 2 minutes
- Close Bypass valve
- Turn on Blower 3
- Wait a little while
Begin Polarization:
- Press "TRANSMIT" on the microwave control panel under Control Unit Status (WARNING: Do not touch the button to the right)
- Press "Take Data" on PDP
- Try to maximize polarization by moving the bellows position
- Watch the frequency in the top left monitor
- You can try to start out in the frequency range in the plot of frequency vs Dose since last annel
- Generally, you have to search for the "best" frequency When searching, make changes by 5MHz, 10MHz at the most
- After making a change, wait about 3 NMR updates to see if it helped/hurt
- Beam trips will change pol quickly and obscure the effect of your changes especially over 50nA
- If the frequencies display does not work look also at the poltarcama/Jpeg/CamImgs.jpg using Camera Set 0, use F5 to update
- As radiation damage accumulates, frequency will move if polarizing positively, frequency will move down and if polarizing negatively, frequency will move up
- Use the logbook − see what frequencies worked for other people
- Log all of your best frequencies when you find them in the Target Operator Microwave Frequency Notebook