Difference between revisions of "Polhe3 meeting 20120418"
From Hall A Wiki
Line 11: | Line 11: | ||
*** if system OK. we will do following (2 weeks):temperature optimization and diffusion test (225~240, may add point 220) | *** if system OK. we will do following (2 weeks):temperature optimization and diffusion test (225~240, may add point 220) | ||
**** temperature stabilized ~ 4 hours | **** temperature stabilized ~ 4 hours | ||
− | **** spin up | + | **** spin up. Note: |
+ | ***** 12sweep * 2 hours ( if spin up time > 6hours), | ||
+ | ***** 15sweep * 1 hours ( if spin up time < 6hours) | ||
**** do EPR | **** do EPR | ||
− | **** do internal temperature test | + | **** do internal temperature test. Note: |
− | **** do AFP loss with laser on, check immediately | + | ***** do NMR laser on - off - on - off- on, each time temperature should be stabilized, ~ 20min), total: 1.5 hour |
− | **** do diffusion model ~ | + | **** do AFP loss with laser on( temperature stabilized) |
− | **** wait for another 3~6 | + | ***** 5~10 sweeps, check immediately,if not stable, should use 10 sweeps. |
− | **** do AFP loss with laser off, check immediately | + | **** do diffusion model ~ 6 hours. Note: |
− | **** do hot spin-down ~ | + | ***** '''it is better to use field sweep for diffusion, for field sweep, you need to changed the phase to get the signal in one axis (X OR Y)''' |
+ | ***** '''at the beginning of diffusion test, do one field sweep- do one frequency sweep - do one field sweep - do one frequency sweep, pay attention to changing phase each time''' | ||
+ | ***** 37 sweep *10 min | ||
+ | **** wait for another 3~6 hour( polarization is become stable),or we can put it in diffusion test. | ||
+ | **** do AFP loss with laser off,( temperature stabilized, 5~ 10 sweeps),check immediately | ||
+ | **** do hot spin-down ~ 18 sweep * 1 hours | ||
**** do one cold spin down for this cell after finishing all temperature tests ~18 hours | **** do one cold spin down for this cell after finishing all temperature tests ~18 hours | ||
Revision as of 12:50, 30 April 2012
- progress
- todo
- Use the Moss cell to check system again.
- if system OK. we will do following (2 weeks):temperature optimization and diffusion test (225~240, may add point 220)
- temperature stabilized ~ 4 hours
- spin up. Note:
- 12sweep * 2 hours ( if spin up time > 6hours),
- 15sweep * 1 hours ( if spin up time < 6hours)
- do EPR
- do internal temperature test. Note:
- do NMR laser on - off - on - off- on, each time temperature should be stabilized, ~ 20min), total: 1.5 hour
- do AFP loss with laser on( temperature stabilized)
- 5~10 sweeps, check immediately,if not stable, should use 10 sweeps.
- do diffusion model ~ 6 hours. Note:
- it is better to use field sweep for diffusion, for field sweep, you need to changed the phase to get the signal in one axis (X OR Y)
- at the beginning of diffusion test, do one field sweep- do one frequency sweep - do one field sweep - do one frequency sweep, pay attention to changing phase each time
- 37 sweep *10 min
- wait for another 3~6 hour( polarization is become stable),or we can put it in diffusion test.
- do AFP loss with laser off,( temperature stabilized, 5~ 10 sweeps),check immediately
- do hot spin-down ~ 18 sweep * 1 hours
- do one cold spin down for this cell after finishing all temperature tests ~18 hours
- Calendar
timeline | Tasks | To be done by | status | |
---|---|---|---|---|
new cell | ||||
04/18-05/02 | receiving from UVa | Jie and Zhiwen | ||
oven bottom piece | ||||
04/18-05/02 | design and make oven bottom piece | Zhiwen and Yawei | ||
04/24-04/27 | set up gradient coil and test | Yawei and Jie | ||
diffusion model measurement | ||||
04/27-05/09 | do the test | all | ||
Dominic test | ||||
05/09-05/19 | set up the comet laser and do the test | all | ||
pulse NMR | ||||
05/19-? | pNMR setup and test | Jie |