Jump to: navigation, search

This is a collection of he3 target calibration measurement for experiment transversity

Please add an entry for each furture calibration. Please include

  • Time stamp as second order title
  • the data directory
  • Result if there is any
  • measurement condition
  • Link to Halog Entry
  • Recommend to include the measurement time line into Halog entry, NOT this WIKI


Contents

Cell History

ASTRAL: Oct 20 - Nov 25, 2008

  • Cell Name: ASTRAL
  • He3 Density (amg): 8.08
  • N2 Density (amg): 0.11
  • Vpc (cc): 164.92
  • Vtt (cc): 6.77
  • Vtc (cc): 79.47
  • NMR Lock-in gain: x10
  • Data : click me

MAUREEN: Nov 26 - Dec 24 2008

  • Cell Name: MAUREEN
  • He3 Density (amg): 7.23 (7.52)
  • N2 Density (amg): 0.102 (0.106)
  • Vpc (cc): 180.75
  • Vtt (cc): 4.16
  • Vtc (cc): 89.05
  • NMR Lock-in gain: x5
  • Data : click me

BRADY: Jan 7 - Feb 4 2009

  • Cell Name: BRADY
  • He3 Density (amg): 7.87
  • N2 Density (amg): 0.11
  • Vpc (cc): 169.27
  • Vtt (cc): 5.98
  • Vtc (cc): 74.57
  • NMR Lock-in gain: x5
  • Data : click me

EPR Calibration

Brady: Transverse Pumping, L-HRS: Negative

Feb 3, 2009

  • Cell: Brady
  • Holding Field: Transverse - (LHRS:- RHRS:0 )
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25x2 + FAP 25)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090203_0944
  • Halog Entry: Click me
  • Calibration Constant: 2.77%/mV

Feb 3, 2009

  • Cell: Brady
  • Holding Field: Transverse - (LHRS:- RHRS:0 )
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25x2 + FAP 25)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090203_0937
  • Halog Entry: Click me
  • Calibration Constant: 2.74%/mV

Jan 28, 2009

  • Cell: Brady
  • Holding Field: Transverse + (LHRS:- RHRS:0 )
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090128_1853
  • Halog Entry: Click me
  • Calibration Constant: 2.81%/mV

Brady: Vertical Pumping, L-HRS: Negative

Feb 5, 2009

  • Cell: Brady
  • Holding Field: Vertical + (LHRS:- RHRS:0 )
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25x2 + FAP 25)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090205_0738
  • Halog Entry: Click me
  • Calibration Constant: 4.47%/mV

Feb 5, 2009

  • Cell: Brady
  • Holding Field: Vertical - (LHRS:- RHRS:0 )
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25x2 + FAP 25)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090205_0730
  • Halog Entry: Click me
  • Calibration Constant: 4.59%/mV


Feb 3, 2009

  • Cell: Brady
  • Holding Field: Vertical + (LHRS:- RHRS:0 )
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx2 + FAP 25W)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090203_1257 
  • Halog Entry: Click me
  • Calibration Constant: 4.55%/mV

Feb 3, 2009

  • Cell: Brady
  • Holding Field: Vertical + (LHRS:- RHRS:0 )
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx2 + FAP 25W)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090203_1246 
  • Halog Entry: Click me
  • Calibration Constant: 4.44%/mV


Jan 28, 2009

  • Cell: Brady
  • Holding Field: Vertical + (LHRS:- RHRS:0 )
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090128_1655
  • Halog Entry: Click me
  • Calibration Constant: 4.67%/mV

Jan 23, 2009

  • Cell: Brady
  • Holding Field: Vertical + (LHRS:- RHRS:0 )
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090123_1514
  • Halog Entry: Click me
  • Calibration Constant: 4.73%/mV

Jan 23, 2009

  • Cell: Brady
  • Holding Field: Vertical + (LHRS:- RHRS:0 )
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090123_1503
  • Halog Entry: Click me
  • Calibration Constant: 4.58%/mV

Brady: Vertical Pumping, L-HRS: Positive

Jan 20, 2009

  • Cell: Brady
  • Holding Field: Vertical + (LHRS:+ RHRS:0 )
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090120_1141 
  • Halog Entry: Click me
  • Calibration Constant: 4.45%/mV

Jan 18, 2009

  • Cell: Brady
  • Holding Field: Vertical + (LHRS:+ RHRS:0 )
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090118_1744
  • Halog Entry: Click me
  • Calibration Constant: 4.53%/mV

Brady: Transverse Pumping, L-HRS: Positive

Jan 12, 2009

  • Cell: Brady
  • Holding Field: Transverse - (LHRS:+ RHRS:0 )
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20090112_1754
  • Halog Entry: Click me
  • Calibration Constant: 2.83%/mV

Jan 11, 2009

  • Cell: Brady
  • Holding Field: Transverse - (LHRS:+ RHRS:0 )
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\Happy Valentine\EPR\Data\EPR_AFP_20090111_1507
  • Halog Entry: Click me
  • Calibration Constant: 2.79%/mV

Maureen: Vertical Pumping, L-HRS: Positive

Dec 22, 2008

  • Cell: Maureen
  • Holding Field: Vertical - (LHRS:+ RHRS:0 )
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081222_1203
  • Halog Entry: Click me
  • Calibration Constant: 5.16%/mV

Dec 22, 2008

  • Cell: Maureen
  • Holding Field: Vertical - (LHRS:+ RHRS:0 )
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081222_1117
  • Halog Entry: Click me
  • Calibration Constant: 5.03%/mV

Dec 16, 2008

  • Cell: Maureen
  • Holding Field: Vertical - (LHRS:+ RHRS:0 )
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081216_1002
  • Halog Entry: Click me
  • Calibration Constant: 5.10%/mV

Dec 13, 2008

  • Cell: Maureen
  • Holding Field: Vertical - (LHRS: Positive polarity)
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081213_1807
  • Halog Entry: Click me
  • Calibration Constant: 5.07%/mV

Maureen: Transverse Pumping, L-HRS: Positive

Dec 13, 2008

  • Cell: Maureen
  • Holding Field: Transverse - (LHRS: Positive polarity)
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081213_0859
  • Halog Entry: Click me
  • Calibration Constant: 2.98%/mV

Dec 5, 2008

  • Cell: Maureen
  • Holding Field: Transverse - (LHRS: Positive polarity)
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081208_1509
  • Halog Entry: Click me
  • Calibration Constant: 2.99%/mV

Maureen: Transverse Pumping, L-HRS: Negative

Dec 5, 2008

  • Cell: Maureen
  • Holding Field: Transverse - (LHRS: Negative polarity)
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081205_0903
  • Halog Entry: Click me
  • Calibration Constant: 2.92%/mV

Dec 2, 2008

  • Cell: Maureen
  • Holding Field: Transverse -
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081202_0833
  • Halog Entry: Click me
  • Calibration Constant: 2.92%/mV

Nov 28, 2008

  • Cell: Maureen
  • Holding Field: Transverse -
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081128_1520
  • Halog Entry: Click me
  • Calibration Constant: 2.92%/mV

Maureen: Vertical Pumping, L-HRS: Negative

Nov 28, 2008

  • Cell: Maureen
  • Holding Field: Vertical +
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081128_1008
  • Halog Entry: Click me
  • Calibration Constant: 5.08%/mV

Nov 26, 2008

  • Cell: Maureen
  • Holding Field: Vertical +
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081126_1028
  • Halog Entry: Click me
  • Calibration Constant: 5.24%/mV

Nov 26, 2008

  • Cell: Maureen
  • Holding Field: Vertical +
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081126_1016
  • Halog Entry: Click me
  • Calibration Constant: 5.23%/mV

Astral: Vertical Pumping, L-HRS: Negative

Nov 25, 2008

  • Cell: Astral
  • Holding Field: Vertical +
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081125_0846
  • Halog Entry: Click me
  • Calibration Constant: 2.48%/mV

Nov 20, 2008

  • Cell: Astral
  • Holding Field: Vertical +
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081120_1145
  • Halog Entry: Click me
  • Calibration Constant: 2.43%/mV

Nov 18, 2008

  • Cell: Astral
  • Holding Field: Vertical +
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081118_1859
  • Halog Entry: Click me
  • Calibration Constant: 2.40%/mV

Astral: Transverse Pumping, L-HRS: Negative

Nov 18, 2008

  • Cell: Astral
  • Holding Field: Transverse - (Correction Coil ON: Small 0.18A)
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081118_1744
  • Halog Entry: Click me
  • Calibration Constant: 1.44%/mV

Nov 18, 2008

  • Cell: Astral
  • Holding Field: Transverse - (Correction Coil ON: Small 0.18A)
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081118_1730
  • Halog Entry: Click me
  • Calibration Constant: 1.45%/mV

Nov 18, 2008

  • Cell: Astral
  • Holding Field: Transverse -
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081118_1711
  • Halog Entry: Click me
  • Calibration Constant: 1.44%/mV

Nov 18, 2008

  • Cell: Astral
  • Holding Field: Transverse -
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081118_1650
  • Halog Entry: Click me
  • Calibration Constant: 1.44%/mV

Nov 15, 2008

  • Cell: Astral
  • Holding Field: Transverse -
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081115_1427
  • Halog Entry: Click me
  • Calibration Constant: 1.48%/mV

Nov 14, 2008

  • Cell: Astral
  • Holding Field: Transverse -
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081114_1704
  • Halog Entry: Click me
  • Calibration Constant: 1.46%/mV

Nov 13, 2008

  • Cell: Astral
  • Holding Field: Transverse -
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx2+FAP 25Wx1)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081113_1121
  • Halog Entry: [Click me
  • Calibration Constant: 1.48%/mV

Nov 11, 2008

  • Cell: Astral
  • Holding Field: Transverse -
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 100W (FAP 20Wx5)
  • Data Directory:
    C:\HappyValentine\EPR\Data\EPR_AFP_20081111_0916
  • Halog Entry: Click me
  • Calibration Constant: 1.45%/mV

Water Calibration

Oct.4, 2008

  • data files:
NMR_Water_20081004_221633
NMR_Water_20081004_221914
NMR_Water_20081004_222533
  • survey photos:
/work/halla/transversity/disk2/Target/WaterNMR/10.05

Oct.11, 2008

  • data files:
NMR_Water_20081011_233531
NMR_Water_20081012_174209
  • survey photos:
/work/halla/transversity/disk2/Target/WaterNMR/10.12

Dec.23, 2008

  • data files:
NMR_Water_20081223_195848
NMR_Water_20081223_200756  

see halog: http://www.jlab.org/~adaq/halog/html/0812_archive/081224154102.html

  • survey photos:
/work/halla/transversity/disk2/Target/WaterNMR/12.24 Xmas Water

Mar.23, 2009

  • data files:
Group 1 (3699 sweeps):
NMR Data: C:\HappyValentine\NMR\Data\NMR_Water_20090320_205008
Field Data: C:\HappyValentine\VI\Field Rotation\Calibration\Field reading
03-20-2009 20-45-33.txt
Group 2 (3699 sweeps):
NMR Data: C:\HappyValentine\NMR\Data\NMR_Water_20090322_111205
Field Data: C:\HappyValentine\VI\Field Rotation\Calibration\Field reading
03-22-2009 11-10-59.txt
see halog: http://www.jlab.org/~adaq/halog/html/0903_archive/090323121956.html
  • survey photos:
/work/halla/transversity/disk2/Target/WaterNMR/other_pic.tar.gz

Internal Temp. Calibration

Feb.3,2009

  • Cell Brady
  • Transverse -

http://www.jlab.org/~adaq/halog/html/0902_archive/090203114627.html



Jan 28,2009

  • Cell Brady
  • Vertical +

http://www.jlab.org/~adaq/halog/html/0901_archive/090128175037.html

Jan 20,2009

  • Cell Brady
  • Vertical +

http://www.jlab.org/~adaq/halog/html/0901_archive/090120130316.html

Jan 17,2009

  • Cell Brady
  • Transverse -
  • Old method

http://www.jlab.org/~adaq/halog/html/0901_archive/090117132035.html

Jan 12,2009

  • Cell Brady
  • Transverse -
  • UVA method from Al

Jan 11, 2009

  • Cell: Brady

http://www.jlab.org/~adaq/halog/html/0901_archive/090111155650.html

Dec 22, 2008

  • Cell: Maureen
  • Oven Temperature: 230C
  • Laser Power: 75 W(Comet 25Wx3) , OFF before calibration
  • Calibrated Field Direction: Vertical +
  • Calibrated NMR RF Field: Longitudinal
  • Calibrated NMR Pumping Chamber Pickup Coil: Transverse
  • DATA ::: C:\HappyValentine\NMR\Data\NMR_He_20081222_221340
  • Time Line:
1:23	Laser Off
1:40	Test Freq Sweep to get field strenth  (since hall truck ramp is 
open). 
	NMR_Freq_He_20081222_134031
	Peak at 84.25kHz
1:43	Freq Sweep NMR
	Data ID : NMR_Freq_He_20081222_134320
	ScreenShotNMR_FrequencySweep_v6.vi_22-12_13-44-24.html
1:45	Frield Rotation to Long. +
1:48	Target -> Pickup Coil
1:56	Field Sweep NMR @ Pick up
	Data ID : NMR_He_20081222_135619
	NMRField_22-12_13-56-54.html
1:56	Move Target Back to He3
1:58	Frield Rotation to Vert. +
2:09	Freq Sweep
	Data ID : NMR_Freq_He_20081222_140851
	ScreenShotNMR_FrequencySweep_v6.vi_22-12_14-11-27.html
2:10	Laser Back

Dec 13, 2008

  • Cell: Maureen

http://www.jlab.org/~adaq/halog/html/0812_archive/081214075556.html

Nov 20, 2008

  • Cell: Astral

http://www.jlab.org/~adaq/halog/html/0811_archive/081121010827.html

Nov 14, 2008

  • Cell: Astral

http://www.jlab.org/~adaq/halog/html/0811_archive/081114174438.html

Nov 5, 2008

  • Cell: Astral

http://www.jlab.org/~adaq/halog/html/0811_archive/081105232737.html

(note::::: This test was performed with the FAP lasers. The first two ( i.e. on Nov. 20 and on Nov. 14 ) tests were performed with COMET lasers)

http://www.jlab.org/~adaq/halog/html/0901_archive/090112193858.html

NMR Cross Calibration

Jan 28, 2009

  • Cell: Brady
  • Holding Field: Vertical +
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Time Line
18:01  NMR Freq Sweep @ Trans
       Data ID : NMR_Freq_He_20090128_180134
18:01  Field -> Long. +; Target -> Pick up
18:07  NMR Field Sweep @ Pickup
       Data ID : NMR_He_20090128_180717
18:07  Field -> Trans -; Target -> he3
18:13  NMR Freq Sweep @ Trans
       Data ID : NMR_Freq_He_20090128_181329
  • Data Dir
C:\HappyValentine\NMR\Data\NMR_Freq_He_20090128_180134
C:\HappyValentine\NMR\Data\NMR_He_20090128_180717
C:\HappyValentine\NMR\Data\NMR_Freq_He_20090128_181329

Jan 20, 2009

  • Cell: Brady
  • Holding Field: Vertical +
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20090120_115442 
C:\HappyValentine\NMR\Data\NMR_He_200900120_120441 
C:\HappyValentine\NMR\Data\NMR_Freq_He_20090120_121308

Jan 18, 2009

  • Cell: Brady
  • Holding Field: Transverse -
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20090118_104320
C:\HappyValentine\NMR\Data\NMR_He_20090118_104948
C:\HappyValentine\NMR\Data\NMR_Freq_He_20090118_105608

Jan 11, 2009

  • Cell: Brady
  • Holding Field: Transverse -
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20090111_170004
C:\HappyValentine\NMR\Data\NMR_He_20090111_170721
C:\HappyValentine\NMR\Data\NMR_Freq_He_20090111_171422

Dec 13, 2008

  • Cell: Maureen
  • Holding Field: Transverse -
  • Oven Temperature: 230C
  • details in the halog entry
  • Data Directory:
C:\Happy Valentine\NMR\Data\NMR_Freq_He_20081213_100659 
C:\Happy Valentine\NMR\Data\NMR_He_20081213_102104
C:\Happy Valentine\NMR\Data\NMR_Freq_He_20081213_103239 

Nov 28, 2008

  • Cell: Maureen
  • Holding Field: Vertical +
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081128_103303
C:\HappyValentine\NMR\Data\NMR_He_20081128_104433
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081128_105403

Nov 25, 2008

  • Cell: Astral
  • Holding Field: Vertical +
  • Pumping Direction: +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081125_090237
C:\HappyValentine\NMR\Data\NMR_He_20081125_091713
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081125_093210

Nov 20, 2008

  • Cell: Astral
  • Holding Field: Vertical +
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081120_122106
C:\HappyValentine\NMR\Data\NMR_He_20081125_124603
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081125_131112

Nov 14, 2008

  • Cell: Astral
  • Holding Field: Transverse -
  • Pumping Direction: -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Data Directory:
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081114_1254xx
C:\HappyValentine\NMR\Data\NMR_He_20081115_0114xx
C:\HappyValentine\NMR\Data\NMR_Freq_He_20081115_0127xx

Spin Up Test

Nov 14, 2008

  • Cell: Astral
  • Holding Field: Transverse -
  • Pumping Direction: -
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\NMR\Data\NMR_Freq_He_20081114_034713
  • Halog Entry: Click me
  • Spin Up Time: 4.65 +- 0.07 hours
  • Max NMR Value: 48.20 +- 0.21 mV

Nov 14, 2008

  • Cell: Astral
  • Holding Field: Transverse -
  • Pumping Direction: +
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\NMR\Data\NMR_Freq_He_20081114_210850
  • Halog Entry: Click me
  • Spin Up Time: 5.03 +- 0.28 hours
  • Max NMR Value: 48.65 +- 0.55 mV

Jan 11, 2009

  • Cell: Brady
  • Holding Field: Transverse -
  • Pumping Direction: -
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    C:\HappyValentine\NMR\Data\NMR_He_20090110_161254
  • Halog Entry: Click me
  • Spin Up Time: 3.82 +- 0.11 hour
  • Max NMR Value: 17.32 +- 0.08 mV = 65.7%

AFP Loss Test

Nov 26, 2008

  • Cell: Maureen
  • Holding Field: Vertical +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    Auto spin flip, sequence 259, flip count# 4-13
  • Halog Entry: Click me
  • AFP Loss per Flip: 0.50%

Nov 25, 2008

  • Cell: Astral
  • Holding Field: Vertical +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    Auto spin flip, sequence 253, flip count# 411-420
  • Halog Entry: Click me
  • AFP Loss per Flip: 0.65%


Nov 20, 2008

  • Cell: Astral
  • Holding Field: Vertical +
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Transverse
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    Auto spin flip, sequence 253, flip count# 72-81
  • Halog Entry: Click me
  • AFP Loss per Flip: 0.50%

Nov 18, 2008

  • Cell: Astral
  • Holding Field: Transverse -
  • NMR RF Coil: Longitudinal
  • NMR Pumping Chamber Pickup Coil: Vertical
  • Oven Temperature: 230C
  • Laser Power: 75W (COMET 25Wx3)
  • Data Directory:
    Auto spin flip, sequence 248, flip count# 9-18
  • Halog Entry: Click me
  • AFP Loss per Flip: 0.54%