    User name R. Michaels

    Log entry time 15:28:07 on February20,2014

    Entry number 385978

    This entry is a followup to: 385972

    Followups:

    keyword=HV modules ok for lower crate

    I tested the HV modules that were in the lower crate. They all work fine in the HV crate on the floor. Below is a snapshot at a high voltage. Some lower voltages were checked with a voltmeter, too.

    Slots 2 and 4 are 1461N. Slot 8 is a 1461P. Slot 12 is 1471N (was being used by VDC).

    Actually, the failure symptoms in lower crate were different.
    The two 1461 would only go between 0 and -34V. If you ask for -50V they'd stop at -34.
    The 1461P would not operate.
    The 1471N would start to ramp up and trip (all channels, even if VDC unplugged).
    It looks like a crate/PS problem. I will talk to Jack and Chuck



    FIGURE 1